Manage the data log

To select data to view, delete or export, specify data filters
1. From the home screen, select Data log.
2. Select an option, then push Select.
Option
Description
View data log
Views measurement data. Select individual lines of data to view more content.
Export data log Exports measurement data from the system to an external device. Preview data selection
before it is exported. Make sure that an external device is connected to the instrument
(e.g., a USB key, external hard drive, etc.).
Delete data log Removes measurement data from the system. Previews data selection before it is
removed.
3. Specify the data selection parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Result type
Application
Date
Operator
4. If View data log was the selected option, push View to see the selected data.
• Push the up and down arrow keys to select a line of data and push Detail to see more data
• If only one Application is selected push the up and down arrow keys to select a line of data
and push Detail to see more data, or push the left and right arrow keys to see related graphs
5. If Export data log or Delete data log was the selected option, push Preview to see the selected
data, then push Export or Delete to start the procedure.
Purge
Use this procedure to remove air bubbles from the system. Refer to
measurement
on page 19 for instructions.
Cell management
Use this option to fill or empty the measurement cell with the pumps. These procedures stop
automatically after a pre-determined time. The time remaining is shown against Security time on the
display.
1. From the home screen, select Cell management.
2. Push the left and right arrow keys to make a selection.
Option
Fill cell (pump 1)
Empty cell (pump
2)
24 English
Description
Sets the type of result available.
Sets the available applications.
Sets the date range.
Sets the available operators.
Description
Use this option to fill the measurement cell with reagent from pump 1.
Chemical exposure hazard. Before selecting this option, examine the level of liquid in
the waste bottle. Make sure that there is enough space in the bottle for all of the liquid
from the measurement cell.
Use this option to empty the measurement cell with pump 2.