13. If the measured values are within the required tolerances, the Passes Test icon will be
displayed. Press the button adjacent to the Forwards Arrow icon, to continue, or press the
button adjacent to the Cross icon to abort the calibration process.
If the measured values are not within the required tolerances, the Fails Test icon is shown with
an error number. Refer to section 11 for a list of error codes.
Press the button adjacent to the Circular Arrow icon to repeat the 2nd calibration solution
measurement or press the button adjacent to the Back Arrow icon to return to the 1st calibration
solution measurement screen.
If the forwards arrow is pressed the instrument compares the two sets of calibration solutions
13. If the measured values are within the required tolerances, the Passes Test icon will be displayed.
which have just been measured and ensures that the values are within 4% (+/-2%) of each
Press the button adjacent to the Forwards Arrow icon, to continue, or press the button adjacent to
other. If the two sets of measured values are not within this tolerance the Fails Test icon is
the Cross icon to abort the calibration process.
shown with an error number. Refer to section 11 for a list of error codes.
14. The instrument will then display the pathlength factors for both 0.2mm and 0.5mm pathlengths.
If the measured values are not within the required tolerances, the Fails Test icon is shown with an
error number. Refer to section 11 for a list of error codes.
Press the button adjacent to the Circular Arrow icon to repeat the 2nd calibration solution
measurement or press the button adjacent to the Back Arrow icon to return to the 1st calibration
solution measurement screen.
If the forwards arrow is pressed the instrument compares the two sets of calibration solutions
Press the button adjacent to the Circular Arrow icon to repeat the 2nd calibration solution
which have just been measured and ensures that the values are within 4% (+/-2%) of each
measurement or press the button adjacent to the Back Arrow icon to return to the 1st calibration
other. If the two sets of measured values are not within this tolerance the Fails Test icon is
solution measurement screen.
shown with an error number. Refer to section 11 for a list of error codes.
15. The new calibration factor values must now be verified by re-measuring the air, blank and
standard values.
14. The instrument will then display the pathlength factors for both 0.2mm and 0.5mm pathlengths.
12. The unit takes three readings at 260 and 330nm,
10. The Calibration Solution Measurement icon will
at both 0.5mm and 0.2mm path lengths. At the end of
be displayed for a second time. Open the instrument
the readings, the unit will display the results of the
lid and use a lint free cloth to clean the read heads.
measurements taken:
11. Pipette a 2.0μl aliquot of the calibration solution onto
the lower read head. Close the instrument lid and
press the key adjacent to the Tick icon to initiate the
calibration solution measurement.
12. The unit takes three readings at 260 and 330nm,
at both 0.5mm and 0.2mm path lengths. At the end
of the readings, the unit will display the results of the
measurements taken:
If the values displayed are within the tolerance range,
the Passes Test icon will be displayed. Press the button
adjacent to the Forwards Arrow icon, to continue, or
press the button adjacent to the Cross icon to abort the
calibration process.
If the measured values are not within the required
tolerances, the Fails Test icon is shown with an error
number. Refer to section 11 for a list of error codes.
16. The Air Measurement icon will be displayed. Open
the instrument lid and use a lint free cloth to clean the
read heads. Close the instrument lid and press the key
If the values displayed are within the tolerance
adjacent to the Tick icon to initiate the blank
range, the Passes Test icon will be displayed. Press
measurement.
the button adjacent to the Forwards Arrow icon, to
17. The unit takes two dark and light readings at 260
continue, or press the button adjacent to the Cross
and 330nm, at both 0.5mm and 0.2mm path lengths.
icon to abort the calibration process.
18. The Blank Solution Measurement icon will be
displayed. Open the instrument lid and pipette a 2.0μl
If the measured values are not within the required
aliquot of the matrix blank solution onto the lower read
head. Close the instrument lid and press the key
tolerances, the Fails Test icon is shown with an error
adjacent to the Tick icon to initiate the blank
number. Refer to section 11 for a list of error codes.
measurement.
23