Dräger Fabius Manual Del Usuario página 11

Tabla de contenido

Publicidad

Reprocessing list
Applicable to non-infectious patients.
The reprocessing list contains guiding values only.
The instructions of the hospital's infection control
officer shall prevail and must be observed by the
user!
Components which
Recommended
can be reprocessed
reprocessing
intervals
Rigid arm for
Weekly
breathing bag
Before using on patients again
1 Gather the necessary device components, see
chapter on assembly and preparation.
2 Mount the parts in reverse order from
disassembly.
3 Check readiness for operation, chapter Getting
Started.
Checklist
Prerequisites
Locating leakages in the breathing circuit
The check must be performed once without
vaporizer and once with vaporizer. Control dial is in
the zero position.
1)
The D-Vapor
must be switched on for the leakage
test. After the leakage test has ended, switch off the
D-Vapor again.
Supplement Fabius
Manual
Machine
precleaning
cleaning and
thermal
disinfection
Yes
Yes
If the system has leakage (the pressure drops):
1) The D-Vapor may not be used with Fabius MRI in an MR
environment.
Manual disinfection
Surface
Disinfection
disinfection
by immersion
bath
Yes
Yes
Change to Standby mode and press the
soft key for the leakage test.
Follow the instructions on the screen.
Check all plug connections and screw
connections for a tight seat.
Replace missing or damaged seals. If
necessary, contact DrägerService or the
local authorized service partner.
English
Hot steam
sterilization
Yes
11

Publicidad

Tabla de contenido
loading

Tabla de contenido