Description of Equipment or Protective System:
The PMA.Vision VIS1Yab-c is designed for the measurement of weight in the hazardous location.
The PMA.Vision contains the processing circuitry and has a touch enable LCD display with
backlight facility. Power and data to the non-hazardous location are provided via the YCO17-Z Ex-
Link Converter (FM15ATEX0035X). All signal outputs to the Ex-Link Converter are in RS232
communications. The PMA.Vision is housed within non-metallic enclosure. A dedicated potential
equalization connection is provided on both the PMA.Vision and the YCO17-Z Ex-Link Converter.
Power and data connections between the YCO17-Z and the PMA.Vision are made via a multi-
conductor CAT6 cable.
PMA.Vision VIS1Yab-c
a = Accessories: Up to three letters and/or numbers or blank (not critical to safety)
b = Metrology Approval: Up to two letters and/or numbers or blank (not critical to
safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
14
Specific Conditions of Use:
1. The PMA.Vision shall only be used with the YCO17-Z Ex-Link Converter (FM15ATEX0035X).
2. The protective ground connectors on the PMA.Vision and the YCO16-Z Ex-Link Converter shall be
connected together using a potential equalisation conductor with a cross-sectional area of at least
4 mm
2
3. The CAT6 cable used between the PMA.Vision and the YCO17-Z Ex-Link Converter shall be
obtained from Sartorius.
4. The enclosure of the PMA.Vision is non-conducting and may generate an ignition-capable level of
electrostatic charges under certain extreme conditions. The user shall ensure that the equipment
is not installed in a location where it may be subjected to external conditions that might cause a
build-up of electrostatic charges on non-conducting surfaces, additionally; cleaning of the
equipment should be done only with a damp cloth.
15
Essential Health and Safety Requirements:
The relevant EHSRs that have not been addressed by the standards listed in this certificate have been
identified and assessed in the confidential report identified in item 8.
16
Test and Assessment Procedure and Conditions:
This Type Examination Certificate is the result of testing of a sample of the product submitted, in
accordance with the provisions of the relevant specific standard(s), and assessment of supporting
documentation. It does not imply an assessment of the whole production.
Whilst this certificate may be used in support of a manufacturer's claim for CE Marking, FM Approvals
Europe Ltd accepts no responsibility for the compliance of the equipment against all applicable Directives
in all applications.
This Certificate has been issued in accordance with FM Approvals Europe Ltd's ATEX Certification
Scheme.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Europe Ltd. One Georges Quay Plaza, Dublin. Ireland. D02 E440
T: +353 (0) 1761 4200 E-mail:
F ATEX 029 (Mar/2019)
108
Installation Instructions VIS1Y
SCHEDULE
to Type Examination Certificate No. FM15ATEX0033X
.
atex@fmapprovals.com www.fmapprovals.com
Page 2 of 3