10 Device Storage

If the device is not used for an extended period of time, it is recommended that the device be
stored in an area that is clean, organized and compliant with the storage conditions

10.1 Storage and transport conditions

Warning
 Please store or transport the device according to the specified transport and
storage conditions
o
o
o
 Storage and transport equipment in the following environments is prohibited
Inflammable, explosive, difficult to solid place and strong electromagnetic
radiation.

10.2 Preparing for storage

6. Power the device off and remove the syringe
7. Disconnect the power cord and all other cables.
8. Remove the pump from the mounting pole.
9. Clean and disinfect the pump and the drop sensor.
10. Carefully handle the device and store it in an area specially prepared for storage.
Warning
 Suggested to charge the battery once every month for long time storage
10.3 Using after storage
Warning
Before use the device after storage or transport,
1. Connect the device to the AC power source to recharge the battery for about
10 hours without switching the device on.
2. Test the device functioning, and check it for traces of damages, especially after
transport.
Temperature range: : -10°C to +55°C
Pressure range: 860 hPa to 1060 hPa
Humidity range: 20% to 90%, no condensation