4.3.4 Time mode
Information:
In Time Mode:
-
The parameters, infusion duration (TIME) and solution volume (SOL VOL), are
set by the user.
-
The flow rate is determined by TIME and SOL VOL settings.
Press
key till the display shows "MODE: TIME".
Press
or
parameter is selected, it will be highlight displayed,
Use
or
Information:
The infusion time range: 00h01min
"SOL VOL" range: 0.1ml
Warning
Improper parameter settings may cause the set flow rate out of limit and trigger
the "Out of Flow Range" alarm,
When select
Information:
Select
and the parameter was set and effective.
Select
Only occlusion alarming pressure level can be set in time mode interface.
Information:
To set the occlusion alarming pressure level, please refer to Chapter 4.4.3.
key to select the parameter to set ( "Time", "SOL VOL", ) , when the
to set the parameter value as required.
:
or
, press
and press
and press
99h59min.
~
9999.9 ml.
~
will be disappeared.
, user interface will be changed.
, the display changes into time mode interface
, the display changes into rate mode interface.
84