4 Maintenance
CAUTION
▪
Equipment maintenance should only be carried out by a MEDIT employee or a MEDIT-
certified company or personnel.
▪
In general, users are not required to perform maintenance work on the i500 system
besides calibration, cleaning and sterilization. Preventive inspections and other regular
maintenance are not required.
4.1 Calibration
Periodic calibration is required to produce precise 3D models.
You should perform the calibration when:
▪
The quality of the 3D model is not reliable or accurate as compared to the previous
results.
▪
Environmental conditions such as temperature has changed.
▪
Calibration period has expired.
You can set the calibration period in the Menu > Settings > Calibration Period (Days)
The calibration panel is a delicate component. Do not touch the panel directly.
Check the calibration panel if the calibration process is not performed properly.If
the calibration panel is contaminated, please contact your service provider.
We recommend performing the calibration periodically.
You can set the calibration period via Menu > Settings > Calibration Period (Days).
The default calibration period is 14 days.
How to calibrate i500
▪
Turn on i500 and launch the image acquisition software.
▪
Run Calibration Wizard from Menu > Settings > Calibration
▪
Prepare the Calibration Tool and the i500 handpiece.
▪
Turn the dial of the calibration tool to position
▪
Put the Handpiece into the calibration tool.
Click "Next" to start the calibration process.
▪
When the calibration tool is mounted properly in the correct position, the system will
automatically acquire the data at position
▪
When data acquisition is complete at position
▪
Repeat the steps for positions
▪
When data acquisition is complete at the
calculate and show the calibration results.
.
1
.
1
, turn the dial to the next position.
1
~
and the
2
8
LAST
position, the system will automatically
LAST
position.
Intraoral Scanner i500
11