6 Electro – Magnetic Compatibility Information
IEC 60601-1-2 Test Level
± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV,
Electrical fast transient/burst IEC 61000-4-4
Mains power quality should be that of a typical commercial or hospital environment.
IEC 60601-1-2 Test Level
± 2 kV 100 kHz repetition frequency
Surge Line-to-line IEC 61000-4-5
Mains power quality should be that of a typical commercial or hospital environment.
IEC 60601-1-2 Test Level
Surge Line-to-ground IEC 61000-4-5
Mains power quality should be that of a typical commercial or hospital environment.
IEC 60601-1-2 Test Level
± 0.5 kV, ± 1 kV, ± 2 kV
Voltage dips IEC 61000-4-11
Mains power quality should be that of a typical commercial or hospital environment. If
the user of the EUT image intensifier requires continued operation during power mains
interruptions, it is recommended that the EUT image intensifier be powered from an
uninterruptible power supply or a battery.
IEC 60601-1-2 Test Level
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
20
User Manual
± 15 kV air
± 0.5 kV , ± 1 kV
Compliance Level
± 8 kV contact ± 15 kV air
Compliance Level
± 2 kV 100 kHz repetition frequency
Compliance Level
± 1 kV
Compliance Level
± 2 kV
Compliance Level
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°