Measurement Specifications - Ideal SureTest 61-164 Manual De Instrucciones

Ocultar thumbs Ver también para SureTest 61-164:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Measurement Specifications:

All specifications are at 23°C ± 5°C at less than 80% relative humidity.
Accuracy is state as ± ([% of range] + [counts]).
AC converter is true rms sensing.
Measurement
Line Voltage
Peak Line Voltage
Frequency
% Voltage Drop
Voltage Loaded
Neutral-Ground V
Impedance - Hot
Neutral, & Ground
GFCI Trip Time
GFCI Trip Current
EPD Trip Current
Estimated Load on
Line
Ranges
85.0 - 250.0 VAC
121.0 - 354.0 VAC
45.0 - 65.0 Hz
0.1% - 99.9%
10.0 - 132.0 VAC
0.0 - 10.0 VAC
0.00 Ω - 3.00 Ω
> 3 Ω
1mS to 6.500S counter.
6.0 - 9.0 mA
30.0 - 37.0 mA
1.00 – 20.0A
Resolution
0.1V
0.1V
0.1Hz
0.1%
0.1V
0.1V
0.01Ω
1 mS
0.1 mA
0.1 mA
0.1A
14
Accuracy
1.0% ± .2V
1.0% ± .2V
1.0% ± .2Hz
2.5% ± .2%
2.5% ± .2V
2.5% ± .2V
2.5% ± .02Ω
Unspecified.
1.0% ± 2mS
1.0% ± .2mA
1.0% ± .2mA
Unspecified