Auto Restart
Auto Restart is used to turn the auto restart feature on/off. The factory default setting is disabled but
it may have been enabled using the quick start procedure.
CAUTION
Note The chiller will not automatically restart if the Show Quick Start was enabled, see page 4-20.
In this case only the controller is powered.
Line Frequency
Auto Restart
Audible Alarms
RA FAN SPEED MODE
Care Level
SERIAL COMM - DCOM
Auto Restart
Audible Alarms
Auto Restart
Audible Alarms
Auto Restart
Audible Alarms
Audible Alarms
Audible Alarms is used to turn the audible alarm on/off.
Line Frequency
Auto Restart
Audible Alarms
RA FAN SPEED MODE
Care Level
SERIAL COMM - DCOM
Thermo Scientific
If the auto restart is enabled and the chiller shuts down as a result
of a power failure, when power is restored the chiller will automati-
cally restart and operate at the saved values. Consider any possible
risks before enabling this mode of operation.
The color of the box next to Auto Restart indi-
60Hz
cates the status,
screen the auto restart is disabled.
off
To change the status highlight the line. Because the
box is also now highlighted
indicates enabled. In this screen the feature is still
disabled.
Press the enter key to change the status. In this screen
Auto Restart is enabled.
Once the highlight is removed confirm the status
of the Auto Restart feature. In this screen the Auto
Restart feature is enabled.
Use the same procedure described above to enable/
60Hz
disable the alarms.
In this screen the alarms are disabled.
off
is enabled,
is disabled. In this
indicates disabled and
ThermoFlex 4-17
Section 4