Technical Specifications - Beurer BM 45 Instrucciones De Uso

Ocultar thumbs Ver también para BM 45:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 53
• Repairs must only be carried out by Customer Services or
authorised retailers. Before making a claim, please check the
batteries first and replace them if necessary.
• For environmental reasons, do not dispose of the
device in household waste at the end of its service life.
Dispose of the device at a suitable local collection or
recycling point in your country. Dispose of the device in
accordance with EC Directive – WEEE (Waste Electrical and
Electronic Equipment). If you have any questions, please
contact the local authorities responsible for waste disposal.
Disposing of the batteries
• The empty, completely flat batteries must be disposed of
through specially designated collection boxes, recycling
points or electronics retailers. You are legally required to
dispose of the batteries.
• The codes below are printed on batteries
containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.

12. Technical specifications

Device
Model no.
BM 45
Type
M1501
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
Classification
33
Cuff pressure 0 – 300 mmHg,
systolic 60– 250 mmHg,
diastolic 40 –199 mmHg,
Pulse 30 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 160 mm x W 107 mm x H 50 mm
Approx. 408 g
(without batteries, with cuff)
22 to 44 cm
+5 °C to + 40 °C, 15 – 90 % relative air
humidity (non-condensing)
- 25 °C to + 70 °C, ≤93 % relative air
humidity, 700hPa – 1060hPa ambient
pressure
4x 1,5 V
AA batteries
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF (cuff), CLASS II (when using a
mains part)