Chapter 4. Pump System Functions
.1. Run/Stop
Press the Run/Stop pad to start
all pump system functions.
Note: numbers in ( ), refer to diagram on page 12.
.2. Preset pressure
The pressure is displayed in
brackets (eg. <50>) on the LCD
(5) for 2 seconds when one of
the pressure pads is pressed,
then the actual pressure is
displayed (pressure settings are
in levels).
Level 50, equals 5 feet
of gravity pressure or the
equivalent of 112.5 mmHg
(±10%).
The pump will maintain the
pre-set level. When operating,
use the �+� and �–� pads (9) to
increase and decrease the level.
The factory preset level is 50.
2
.. Filling the chamber
Press the Fill Chamber pad
to fill the pressure chamber
initially, and throughout
the surgery as needed. The
chamber should never be over
1/3 full.