Annual Inspection - DePuy Mitek fms solo Manual De Instrucciones

Ocultar thumbs Ver también para fms solo:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Weekly inspection:
FMS recommends a weekly inspection of all
equipment. These inspections can be performed on
a less regular basis depending on the duration and
frequency of use.
- Perform a visual inspection of the device. If
damage is noted return to FMS Service Centre for
internal damage inspection.
- Inspect that the transparent safety cover (1) and
black hinges are in good working order.
- Inspect power cord and all cables for wear
or cuts.
- Ensure that the foot pedal and shaver interface
connector pins are not missing or bent.
- Plug the power cord to the power connection.
- Switch the I/O switch (14) on �I�.
- Turn the device ON by pressing the POWER
(4) pad.
- Press the Pump Run/Stop pads (10).
- Press on all pads (an audible tone should be
heard for each pad pressed).
) on �I�.
POWER


Annual inspection:

FMS recommends that a bio-medical engineer
or authorized service personal annually check
all equipment to assess its functionality and
technical safety.
Refer to Service manual for annual
inspection procedures.