1. Before operating the equipment, verify the following:
▪ You are using Stryker-approved equipment.
▪ The equipment is not damaged.
▪ The equipment is clean and sterile.
▪ You have properly assembled the equipment.
▪ You have installed a fully-charged power module.
NOTE: If you experience difficulty during operation,
contact your Stryker sales representative or call Stryker
customer service. See the Contact Information section.
Outside the US, contact your nearest Stryker subsidiary.
2. Set the safety switch to the run mode position.
NOTE: Set the safety switch back to the safe mode
position when the handpiece is idle or when passing the
handpiece to another person.
CAUTION: DO NOT apply excessive pressure or restrict
the motion of the cutting tool during operation. An
excessive load on the system may cause the power
module to temporarily shut down. If this happens, release
the trigger to resume operation or allow the power module
to cool for 20 to 40 minutes.
3. Pull the trigger to control the variable speed operation
of the handpiece.
NOTE: Typical power module run time is 5 minutes or
longer, depending on surgical procedure and technique.
8

To Remove a Power Module

NOTE: Keep the power module as clean as possible
because it cannot be sterilized.
1. Wipe all debris from the external surfaces of the
handpiece.
2. Turn the latch to the unlock position and open the
door.
3. Use the handle to remove the power module from the
handpiece.
7306-001-700 Rev-A
www.stryker.com