X - ELECTROMAGNETIC COMPATIBILITY
Important:
The power cable, the handpiece cable and the footswitch cable must be kept apart.
The P5 N
XS requires special precautions to be taken with regard to electromagnetic
EWTRON
compatibility.
It must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the P5 N
EWTRON
The recommended separation distances in this paragraph must therefore be complied with.
The P5 N
XS must not be used near or on top of another device.
EWTRON
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have
the consequence of increasing the emissions or decreasing the immunity of the P5 N
10. 1 ELECTROMAGNETIC EMISSIONS
The P5 N
XS is intended for use in the electromagnetic environment specified in the table
EWTRON
below.
The user and/or installer must ensure that the P5 N
Emission test
RF emission - CISPR 11.
RF emission - CISPR 11.
Harmonic current emission
IEC 61000-3-2.
Voltage fluctuation and flicker
IEC 61000-3-3.
16
XS.
Compliance
Group 1
Class B
Class A
Complies
XS is used in such an environment.
EWTRON
Electromagnetic environment - guidance
P5 N
XS uses RF energy for internal operation.
EWTRON
Therefore, its radiofrequency emissions are very low and are not likely
to cause any interference in nearby equipment.
P5 N
XS is suitable for use in all establishments, including
EWTRON
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
XS.
EWTRON