d) Left-hand side of the control unit
The knob (Fig. 2-3) governs:
- Switching off the irrigation function when
turned completely counterclockwise.
- Irrigation flow rate adjustment: from "min" to
"max".
F
/P
function, when pressed. Press this
LUSH
RIME
button to start draining with a delay of 8 min
(time required to empty a full tank); press again
to stop the draining and delay.
Note:
• Knob in position
around 5ml/min.
• Knob in position
MAX
around 40ml/min.
• F
/P
function corresponds to a flow of
LUSH
RIME
around 40ml/min.
e) Technical characteristics
Device name: P5 N
EWTRON
Electrical power supply:
Voltage: 100VAC / 115VAC / 230VAC (factory set)
Frequency: 50Hz/60Hz
Rated power: 50VA @ 230VAC
Operating:
Intermittent service: 10min ON / 5min OFF
Output characteristics:
Vibration min. frequency: 28kHz
Irrigation flow rate: 0 to 40ml/min (nominal
Protection:
Electrical safety: Class 1, type BF
2 fuses (power receptacle):
5 mm x 20 mm / 315mAT @ 230 V rating 35A
5 mm x 20 mm / 400mATT @ 115 VAC rating 35A
5 mm x 20 mm / 400mATT @ 100 VAC rating 35A
1 FU3 fuse : CMS / 2AT
Overall dimensions:
Control unit:
Width: 158mm
Height: 88mm
8
corresponds to a flow of
MIN
corresponds to a flow of
XS
value)
Height with tank: 125mm
Depth: 290mm
Weight: 1700 g
Footswitch:
Width: 70mm
Height: 30mm
Depth: 95mm
Weight: 150g
Handpiece cable: 2.000mm
Tank capacity: 300ml
Temperatures:
Operating: +10°C to +40°C
Storage: -20°C to +70°C
Humidity:
Operating: 30% to 75%
Storage: 10% to 100% including condensation
IV - INSTALLATION / FIRST
USE
4. 1 UNPACKING THE DEVICE
Upon receipt of the device, check for any damage
caused in transit.
Contact your supplier if necessary.
4. 2 RECOMMENDATIONS
Have your device connected to the electric
supply by a certified dental installation
technician.
P5 N
XS electrical connections must be in
EWTRON
compliance with current standards. A grounded
power socket must be used.