The P M
N
AX
EWTRON
following items on the upper side:
- a control dial (Fig. 1-4), with numbers ranging
from 1 to 20 (power level) (Fig. 1-2).
- a knob for adjusting the irrigation solution flow
and the Flush/Prime function (Fig. 1-6).
- an irrigation solution tank selection button (Fig.
1-5).
- an "air" function indicator (Fig. 1-8).
The P M
N
AX
EWTRON
following item on the front panel a handpiece
cord connector (Fig. 1-7).
Located at the back of the unit are:
- a power cord socket with ground pin (Fig. 2-1),
- a footswitch connector (Fig. 2-2),
- an air input connection (Fig. 2-3).
- a power switch (Fig. 2-4).
The handpiece rest is located on the right-hand
side (Fig. 1-9).
The device has ventilation vents on the rear and
lower panels.
The device has four rubber feet to allow correct
air flow.
The handpiece accessories and tips are delivered
with their own instructions.
XS control unit has the
XS control unit has the

3. 2 TECHNICAL DESCRIPTION

a) Control unit upper panel
The dial (Fig. 1-3) is used to adjust the power: 1
to 20.
The display is separated into four colored
sections, each with five positions (Fig. 1-2).
The "ON" indicator (Figure 1, (4)) llights up with
the color and number of the chosen power level:
- Green: 1 to 9: very low to low power (e.g.
periodontics).
- Yellow: 6 to 14: medium
endodontics)..
- Blue: 12 to 18: high power (e.g. prophylaxis).
- Orange: 17 to 20: very high power (e.g.
The B
mode can be used to deliver an
OOST
increased power of 2.5 steps on top of the power
effectively used (e.g.: power 6 is displayed =>
boost = 8.5).
When the B
function is selected, an orange
OOST
indicator lights up on the settings dial.
The technology of this device allows the
practitioner to work at constant amplitude
depending on the power selected with
compensation according to the effort applied.
Note:
The tip usage power level is selected according to
the directions in the TipBook™ and the color-
coded system (CCS) tips.
The Boost function is automatically deactivated
when the practitioner selects a new power
setting with the power setting dial.
The knob on the upper panel (Fig. 1-6) governs:
- Irrigation solution flow adjustment.
- F
/P
function, when pressed. Press this
LUSH
RIME
button to start draining with a delay of 8 min
(time required to empty a full tank); press again
to stop the draining and delay.
power
specific treatments).
(e.g.
7