Note:
• Knob in position
around 5ml/min.
• Knob in position
MAX
around 40ml/min.
• F
/P
function corresponds to a flow of
LUSH
RIME
around 40ml/min.
The push button (Fig. 1-5) on the front panel
governs irrigation solution tank selection.
Note: The selected tank is lit up.
When the indicator (Fig. 1-8) is lit up the "air"
function is activated.
b) Control unit back panel
- The power receptacle (Fig. 2-1) with its
grounding pin is used to connect the P M
N
XS to the power supply using a plug-in
EWTRON
power cord.
- The footswitch connector (Fig. 3-2) is used to
connect the device to the multifunction control
footswitch.
- The device is connected to the dental office's
medical-quality air distribution supply via the
air connection (Fig. 3-3).
- The power switch (Fig. 3-4) turns the device on
or off.
- The fuse box (Fig. 3-5) contains two fuses to
protect the device.
c) Control unit front panel
The handpiece cord connector (Fig. 1-7) is used
to:
- Connect the handpiece cord to the device.
- Circulate fluid to the spray.
- Electrically link the device and the handpiece
via the handpiece cord.
8
corresponds to a flow of
MIN
corresponds to a flow of
d) Control footswitch
The multifunctional footswitch (Fig. 3-1) is used
to:
- Start the P M
(Fig. 3-2).
- Start or stop the irrigation function (Fig. 3-3).
- Select the air function (Fig. 3-4).
- Select the boost function (Fig. 3-5).
Its weight and anti-slip pad ensure good stability.
e) Technical characteristics
Device name: P M
Electrical power supply:
Voltage: 100VAC / 115VAC / 230VAC (factory set)
Frequency: 50Hz / 60Hz.
Rated power: 30VA @ 230VAC.
Operating:
Intermittent service: 10 min ON / 5 min OFF.
AX
Output characteristics:
Vibration frequency: 28KHz to 36Khz.
Irrigation flow rate: 0 to 40ml/min (nominal
Air flow: 1.4l/min to 1.6l/min depending on the
tip.
Protection:
Electrical safety: Class 1, type BF.
2 fuses (power receptacle):
5 mm x 20 mm / 315 mAT rating 35 A @ 230 VAC.
5 mm x 20 mm / 500 mAT rating 35 A @ 115 VAC.
5 mm x 20 mm / 500 mAT rating 35 A @ 100 VAC.
1 FUS1 ref. fuse not accessible to the user: CMS /
2 AT.
N
XS ultrasonic function
AX
EWTRON
N
XS
AX
EWTRON