Errors and solutions
9.
The protective diaphragm effectively prevents that liquids or solids are sucked into the
pump. If substances are sucked in nevertheless or any other problem occurs, immediately
put the device out of operation. Contact your dealer or the customer service of your shop.
Problem
Pump does not start
Pump does not suck
(does not build up
vacuum)
The breast pump/power unit must not be opened by users/patients. The device is
maintenance-free. For repair work, the device must be returned to the manufacturer.
10. Environmental protection and disposal
Please dispose of a disused mamivac
disposal regulations for electrical devices.
11. Technical data
The expected service life is 2 years.
mamivac LACTIVE
Nominal voltage:
Power consumption:
H x W x D (mm):
Weight:
Max. vacuum:
Mean vacuum:
Power supply unit
Nominal input voltage:
Input current:
Output voltage:
Output current:
Protection class:
H x W x D (mm):
Weight:
Error/solution
- Check connections of power supply.
- LED on power unit must light (if existing).
- If the problem is not solved, please contact the customer
service of your shop.
- Switch off and on the device several times.
Wait a minute before switching on again.
- Check the respective suction kit for proper assembly.
- Check tube valve (for cracks and dirt).
- Check protective diaphragm for cracks.
- Protective diaphragm must be inserted flush and crinkle free.
- Check hose connections.
LACTIVE breast pump in accordance with your local
®
REF 281024
12 V DC
1.00 A
115 x 125 x 195
630 g
-33 kPa (± 15%)
Device for intermittent sucking without continuous vacuum
REF 283005
100-240 V AC 50 / 60 Hz
500 mA
12 V DC
1,25 A
II
75 x 46 x 30,5
120 g
Device classification:
Classified acc. to 93/42/EEC:
Article 9, Appendix IX:
Class IIa/Rule 11
EU, UK, US version
- 23 -