Storage
Place the electrode in the storage protector
electrolyte (Crisolyt A, KCl 3M sat. AgCl).
Life of an electrode
The average life expectancy of a pH electrode is approximately one
year depending on the working conditions.
An electrode used for more than six months that does not response
appropriately after regeneration, must be replaced by a new one.
To avoid accidental breakage of the glass body electrodes, the use of
the working protector, code 91 62, is recommended.
If an electrode has too short lifespan, less than three months,
ask us about it. Probably there are more suitable electrode for the
application.
Specifications
pH measuring range
Operating temp.
Reference element
Diaphragms
Electrolyte
Body material
Min. immersion depth
Applications:
• Aqueous samples in general.
Limits
Low-conductivity, dirty or viscous solutions.
Products with colloids or solids in suspension.
Samples containing sulphides, reducing sugars or other substances
which react with the silver ions.
Temperatures > 80 ºC.
Warranty
CRISON guarantees this electrode against manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the electrode.
CERTIFICATE OF QUALITY
The attached electrode 50 50 has passed the quality test and follows
the next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s
0 ... 14
0 ... 80 °C
Ag wire coated with AgCl
ceramic
gel
polycarbonate
18 mm
with appropriate