_______________________________________________________________________________
400kHz
4MHz
Input sensitivity: 2V rms.
Note: When the voltage amplitude of the frequency signal inputted is above 50V AC rms, it is recommended
that users set the range selector knob to the ACV range position, and press HZ/Duty button to get into the
mode of Frequency Measurement to test the frequency.
3-2-7. Temperature
Range
Accuracy
(-2
±(1.0%+5)
(30
±(1.9%+15)
±(1.2%+6)
(-1
(500
±(1.9%+25)
Sensor: K Type banana Plug
3-2-8. Diode and Continuity Test
Range
Reading
Forward voltage drop of Diode
Buzzer makes a long sound, While
resistance is less than (50+20)
Over range protection: 250V rms.
4. Operation Method
4-1. Front Panel Description
(1) Jaw
(2) Trigger
(3) Range selector knob: The selector knob is used to select ACA, ACV, DCV, Resistance, Frequency,
Temperature measurement.
(4) Power button
(5) Hold button: Press the button, the reading is held on LCD; Press the button again, it will release the
hold.
(6) LCD
(7) "Voltage, Resistance, Frequency and Temperature" input terminal.
(8) Common input terminal
(9) Wrist strap
(10) Barrier
4-2 AC Current Measurement
(1) Set the range selector knob to ACA Range position.
(2) Open the jaw and clamp on to one conductor, for getting the most accurate reading, it is recommended
that the conductor is placed in the center of the closed jaw.
Note: Do not input the allowable maximum value, if exceeding, there is the risk damaging the instrument
circuit.
4-3. AC Voltage Measurement
(1) Set the range selector knob to ACA Range position. Insert the red lead into "V. " input terminal and
the black lead into "COM" input terminal.
(2) Connect the test leads crossly with the tested circuit, and have the reading from the LCD.
©Copyright Abacanto Digital SA.
100Hz
1kHz
Over range protection: 250V rms.
Resolution
1
1
1
1
Warning: Do not input any voltage at this range.
Warning: Do not input any voltage at this range
Manual de usuario AD6016A
Testing Condition
Forward DCA is approx. 0.5m A, backward
Voltage is approx. 1.5V
Open voltage is approx. 0.5V