Replacing the rechargeable battery

WARNING
!
Do not remove or replace the rechargeable battery in areas
subject to possible explosion.
1. Remove both screws with an appropriate hex key.
2. Pull the rechargeable battery out of the Dräger X-act 5000 Basic.
3. Insert a charged NiMH battery into the Dräger X-act 5000 Basic.
4. Fasten both screws with an appropriate hex key.
Dräger X-act 5000 Basic
4

Errors and warnings

4.1

Errors

Errors will be displayed together with the corresponding icon and
a code number or a written hint. The red LEDs will flash until
acknowledged with the "OK" key.
Text/Code
Cause
LEAK TEST
SO
3
ERROR
correctly in the
instrument; or air path
in the instrument is
leaky.
TUBE
Inserted tube or
BLOCKED
sampling system is
clogged or air path in
the instrument is
clogged.
BATTERY
Rechargeable battery
ERROR
is deeply discharged or
the type of charger is
not correct.
Rechargeable battery
is defective.
100
ROM CRC fault
00721900.eps
101
RAM Integrity fault
102
Info flash fault
103
Charger
communication fault.
Remedy
filter does not fit
Repeat leak test. If leak test
failed, check the SO
correct seating or contact
DrägerService
Use new tube or sampling
system, check that SO
fits tight or contact
DrägerService.
Wait for 1 minute. Then clear
the error by either pressing the
"OK" key, using the correct
charger or contacting
DrägerService.
Replace rechargeable battery.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Remove and reinstall
rechargeable battery or
contact DrägerService.
Errors and warnings
filter for
3
®
.
filter
3
17