Cleaning; Maintenance Work; Calibrate Flow; Replacing The So 3 Filter - Dräger X-act 5000 Basic Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5.2

Cleaning

The Dräger X-act 5000 Basic can be cleaned using a damp cloth.
Coarse dust particles can be removed with a soft brush.
5.3

Maintenance work

5.3.1

Calibrate flow

NOTICE
i
i
For flow calibration, the Dräger calibration tube X-act 5000
(8103651) must be used. Observe the supplied instructions
for use 9022933.
The calibration will be done using a flow meter.
1. Select "CALIBRATE FLOW" in the password-protected menu.
2. Follow the instructions on the display.
3. If necessary, set the required values using the arrow keys and
acknowledge the settings using the "OK" key.
Dräger X-act 5000 Basic
5.3.2
Replacing the SO
filter
3
1. Loosen both screws. Remove the SO
2. Replace the SO
filter (2).
3
3. Reattach the SO
filter door. Tighten the screws.
3
4. If necessary, enter the date of the filter change in the password-
protected menu "CHANGE SO3 FILTER".
NOTICE
i
i
Once installed, the SO
installation date can be entered under "Password-protected
menu".
6

Storage

Store the Dräger X-act 5000 Basic in a dry place, free of dust. Keep
out of direct sunlight or away from heat radiation.
filter door (1).
3
filter has a service life of 2 years. The
3
2
1
00821900.eps
Storage
19

Publicidad

Tabla de contenido
loading

Tabla de contenido