OHAUS Valor 3000 Serie Manual De Instrucciones página 19

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN-18
7.4 Compliance
Compliance to the following standards is indicated by the corresponding mark on the
product
.
Marking
Important notice for verified weighing instruments
The first stage of the initial verification has been carried out at the manufacturers
work. It comprises all tests according to the adopted European standard EN
45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the
weighing instrument must strictly observe the re-verification period and inform the
respective W+M authorities.
Standard
This product conforms to the EMC directive 2004/108/EC, the Low
Voltage Directive 2006/95/EC and the Non-automatic Weighing
Instruments Directive 2009/23/EC. The complete Declaration of
Conformity is available online at www.ohaus.com/ce.
AS/NZS 61000.6.1, AS/NZS 61000.6.3
NSF/ANSI 169
NSF/ANSI /3-A 14159-1
Weighing Instruments verified at the place of manufacture bear one
of the preceding mark on the packing label and the green 'M'
(metrology) sticker on the descriptive plate. They may be put into
service immediately.
Weighing Instruments to be verified in two stages have no green 'M'
(metrology) on the descriptive plate and bear one of the preceding
identification mark on the packing label. The second stage of the
initial verification must be carried out by the approved service
organization of the authorized representative within the EC or by the
national weight & measures (W+M) authorities.