OPERATING ISSUE
Scale is not communicating
by serial.
No weight is displayed on the
scale.
The scale is showing brackets
at the top of display.
Capacity: 150 x 0.05 lb / 60 x 0.02 kg
lb
kg
The scale is showing brackets
at the bottom of display.
Capacity: 150 x 0.05 lb / 60 x 0.02 kg
lb
kg
If the problem persists, or if the troubleshooting section does not resolve or describe your problem, contact Uline
Customer Service at 1-800-295-5510.
PAGE 6 OF 18
TROUBLESHOOTING
Invalid pairing in the PC RS232
connection setting.
Power supply may be unplugged or
display cable is not attached to scale
base.
Something is on or touching the scale.
Scale is over capacity. Too much
weight is on the scale.
0
Scale is under capacity. Weight is on
the scale at power up.
0
CAUSES
1-800-295-5510
uline.com
RECOMMENDATIONS
Select the correct scale model in the
shipping software ("Toledo," "PS60," or
"BC60").
If using RS232, confirm power supply
is plugged into working AC outlet, USB
cable is plugged into power supply
and USB cable is plugged properly
into scale base.
If using USB, make sure USB cable is
pushed in all the way at both scale
and PC.
Unplug and reinsert the display
connector into the scale base.
Make sure nothing is on or touching
the top of the scale when powering
up.
Reduce weight.
Disconnect power by unplugging the
USB cable or power supply. Remove
any weight from the scale, then plug
the scale back in. The scale will
reboot and capture zero.
Zero scale with no weight on platter.
Disconnect power by unplugging the
USB cable or power supply. Remove
any weight from the scale, then plug
the scale back in. The scale will
reboot and capture zero.