To avoid high backpressure, we recommend rinsing the column after installation at a low flow rate (0.1 mL/min)
for approx. 20 min until the working temperature is reached.
Guard column
Metrosep A Supp 5 Guard/2.0 (6.1006.600)
Metrosep A Supp 5 S-Guard/2.0 (6.1006.610)
Metrosep A Supp 16 Guard/2.0 (6.1031.600)
Metrosep A Supp 16 S-Guard/2.0 (6.1031.610)
Storage
Store the column in eluent at 4 to 8 °C.
Do not store the column below 0 °C.
Regeneration
Ensure that the maximum pressure is never exceeded during regeneration.
If the pressure becomes too high, reduce the flow rate.
Depending on the type of contamination, proceed in accordance with the following instructions.
Contamination with low-valence hydrophilic ions
1. Rinse with ultrapure water (25 min at 0.1 mL/min)
2. Rinse with 10x concentrated standard eluent (100 min at 0.1 mL/min)
3. Rinse with ultrapure water (25 min at 0.1 mL/min)
4. Rinse with eluent (100 min at 0.1 mL/min)
Contamination with high-valence hydrophobic ions or organic contaminations
1. Rinse with ultrapure water (25 min at 0.1 mL/min)
2. Rinse with 100% acetonitrile (20 min at 0.1 mL/min)
3. Rinse with ultrapure water (25 min at 0.1 mL/min)
4. Rinse with 10x concentrated standard eluent (100 min at 0.1 mL/min)
5. Rinse with ultrapure water (25 min at 0.1 mL/min)
6. Rinse with eluent (100 min at 0.1 mL/min)
General notes
Sample solutions must be microfiltered (0.45 µm).
To protect the separation column, we recommend using the pulsation absorber (6.2620.150) to reduce the
injector pressure surges.
To avoid high backpressure when changing from/to organic modifiers, adjust the flow rate in small increments
from 0.1 mL/min to the standard conditions within 30 min.
FR
Matériau de la colonne
Alcool polyvinylique avec groupes d'ammonium quaternaire, diamètre de particule 5 µm
Dimensions
6.1006.640 : 150 x 2,0 mm
6.1006.650 : 250 x 2,0 mm
Gamme de pH
3 à 12
Gamme de température
20 à 60 °C
Température standard : 45 °C
Pression maximale
20 MPa (200 bars)