  • Página 1 CH-9101 Herisau/Suiza E-Mail info@metrohm.com Internet www.metrohm.com 758 KFD Titrino Versión del programa 5.758.0022 Instrucciones para el uso 8.758.1205 11.2001 / pkl...
  • Página 3: Tabla De Contenido

    3.2.2 Description of the remote control commands ........... 99 3.3 Properties of the RS 232 Interface ................130 3.3.1 Handshake....................... 130 3.3.2 Pin Assignment ....................133 3.3.3 ¿Qué hacer si la transmisión de datos no funciona?........135 758 KFD Titrino...
  • Página 4 6.4.3 Mantenimiento y reajuste del Titrino..............194 6.5 Garantía y conformidad..................195 6.5.1 Garantía ......................195 6.4.2 Declaración de conformidad ad UE ..............196 6.4.3 Certificado para la conformidad y validación del sistema....... 197 6.6 Referencias de pedido, accesorios................ 198 Indice ..........................202 758 KFD Titrino...
  • Página 5: Introducción

    1 Introducción 1.1 Descripción del aparato El 758 KFD Titrino es un valorador de titración para determinaciones rápidas y precisas a Karl Fischer y punto final con pantalla LCD gráfica. Las valoraciones se pueden seguir en tiempo real. El 758 KFD Titrino tiene la capacidad de controlar hasta tres dosificadores de valoración.
  • Página 6: Elementos De Manjero

    Introducción 1.2 Elementos de manjero Parte anterior del aparato: 758 KFD Titrino...
  • Página 7 "STATISTICS" La luz está encendida cuando la función "esta- dística" (cálculo de la media y desviación stan- dard) está conectada. "SILO" La luz está encendida cuando la memoria silo (para datos de muestras) está conectada. 758 KFD Titrino...
  • Página 8: Parte Posterior Del Aparato

    Introducción Parte posterior del aparato: 758 KFD Titrino...
  • Página 9 En el caso de redes en las que el voltaje esté sujeto a fuertes interfe- rencias HF, hay que conectar el Titrino por medio de un filtro de la red adicional, p.ej. el modelo Metrohm 615. Conexión para buretas externas D1 y D2 685 Dosimates o 700 Dosinos Conexión para el agitador...
  • Página 10: Manejo Manual

    Borra valores y ajusta valores especia- les. ENTER Acepte valores. STOP Para métodos. QUIT Deje consultas y pausas. START Comienza métodos. Las terceras funciones de las teclas (inscripción superior) del teclado se utilizan para introducir fórmulas, vea página 42. 758 KFD Titrino...
  • Página 11: Principio De La Entrada De Datos

    >aparatos periféricos <QUIT> otra vez deje las consultas "configuration". >ajustes varios >ajustes para RS232-COM1 >ajustes para RS232-COM2 • Cuando puede girar la pantalla (scroll) está marca- >variables comunes do con flechas a la derecha abajo o arriba. >prep.buretas 758 KFD Titrino...
  • Página 12: Entrada De Textos

    <ABC>. 3. Se puede borrar los caracteres de atrás hacia delante con <CLEAR> o añadir más caracte- res. 4. Cuando el texto está completado, deje su en- trada con <QUIT> y acepte el nuevo nombre con <ENTER>. 758 KFD Titrino...
  • Página 13: Configuración, Tecla

    El control se efectúa al final de la titración y tras conectar el Titrino y si hay una calibración actual. Si entra en función de control se emite el mensaje "calibre electrodo". El contador se pone a cero con la próxima calibración. 758 KFD Titrino...
  • Página 14: Ajustes Para Aparatos Periféricos

    Modelos MP8, MC1 Mettler: Modelos AM, PM, AX, MX, UMX y balanzas con interfaces 011, 012 y 016 Mettler AT: Modelo AT AND: Modelos ER-60, 120, 180, 182, FR-200, 300 y FX-200, 300, 320 Precisa: Modelos con interface RS232C 758 KFD Titrino...
  • Página 15 No es aconsejable en trabajos con un Cambiador de muestras. Tiempo de espera (0...999999 s) t(espera) Tiempo de espera después del comienzo antes de que el método empiece. El tiempo de espera se puede interrumpir con <QUIT>. 758 KFD Titrino...
  • Página 16 Impresión sobre "prep" (sí, no) impresión: Si la función está conectada, se imprime automáticamente un informe después de prep. El informe se emite por el mismo COM como los informes desencadenados de manera manual (vea página 10). 758 KFD Titrino...
  • Página 17: Parámetros Para Dosinos

    Longitud del tubo de aspiración (1...999.9 cm) long.tubo asp. 25.0 cm Diámetro del tubo de aspiración (0.1...9.9 mm) diám.tubo asp. 2.0 mm Velocidad de dosificación (0.01...150 mL/min, max.) veloc.dos.DX max. ml/min Velocidad de llenado (0.01...150 mL/min, max.) veloc.rell.DXmax. ml/min 758 KFD Titrino...
  • Página 18 Bloquear >bloquear "sí" significa que la función correspondiente no es accesible. <configuration>: La tecla correspondiente está bloqueada. <parameters>: <smpl data>: La función correspondiente de la memoria interna del cargar método: Titrino está bloqueada. memorizar método: borrar método: 758 KFD Titrino...
  • Página 19 1 es la mayor anchura. En este caso se puede perder la escritura del eje en la margen derecha. Longitud (0.01...1.00) longitud 0.05 Eje del tiempo: Longitud de la curva 0.01 100 cm 10 cm 2 cm 1 cm 758 KFD Titrino...
  • Página 20: Selección Del Modo, Tecla

    TIP es una "envoltura vacía" para la que hay que definir el desarrollo de la secuencia con la tecla <DEF>, vea página 52. Se encuentra una vista general de los modos de titración en las instrucciones breves. 758 KFD Titrino...
  • Página 21: Parámetros, Tecla

    Gama de regulación (la gama de entrada depende de gama regulación 100 mV la cantidad medida: **titr. U, en Ipol: 1..2000 mV I, en Upol: 0.1...200.0 µ Fuera de la gama de regulación se dosifica continuamente, vea página 23. 758 KFD Titrino...
  • Página 22: Parámetros De Titración

    – : En el sentido de voltaje bajo, pequeñas corrientes. Pausa 1 (0...999999 s) pausa 1 Tiempo de espera, por ejemplo, para el equilibrado del **titr. electrodo después del Start. Se puede interrumpir la pausa con <QUIT>. 758 KFD Titrino...
  • Página 23 "no" significa que la prueba no se realiza. Temperatura de la titración (-170.0...500.0 °C) temperatura 25.0 °C Con una termosonda, se mide la temperatura después cond. del comienzo de la titración. 758 KFD Titrino...
  • Página 24 Corrección de la deriva (auto, man. no) corr.deriva: auto: El valor de la deriva está tomado cond. automáticamente en el momento del comienzo. Valor de la deriva para la corrección de la deriva manual valor deriva 0.0 µl/min (0...99.9 L/min) µ cond. 758 KFD Titrino...
  • Página 25 Ajústelo a "no" si no ha conectado un horno al Titrino o si no lo ha hecho con un RS232. Salida del pulso en la línea "activate" (L6, pin 1) del puls.activación: clavijero "Remote" (prim., todo, acond., no) cond. vea página 174. 758 KFD Titrino...
  • Página 26 Se desconecta el agitador. (Desconectar agitador) Se efectúan el cálculo. Cálculos Se emiten los datos. Salida de datos (Reacondicionamiento) Después de la titración se procede al nuevo acondicio- namiento (si está seleccionado). 758 KFD Titrino...
  • Página 27 • Las determinaciones del título deberían realizarse no sólo en el mismo modo como para las muestras sino también con los mismos parámetros. • Los parámetros estándares se recomiendan para la mayoría de titraciones KFT. • Con Ipol resultan curvas más pendientes que con Upol. 758 KFD Titrino...
  • Página 28 I o U deben estar provistos con el prefijo negativo ya que, de lo contrario, sólo resultan unas combinaciones de valores sin sentido alguno. • Para el modo KFT Upol rigen, analógicamente, las mismas leyes que para KFT Ipol. 758 KFD Titrino...
  • Página 29: Muestras Sólidas

    Si tiene que abrir la célula de titración KF a cada adición de la muestra, determine un valor en blanco y considérelo en el cálculo del resultado (vea página 180). 758 KFD Titrino...
  • Página 30 El tiempo de las titraciones es • Con reactivos de 2 componentes se puede haber cada vez más largo. agotado la capacidad del tampón del disolvente. Cambie la solución. • Si aumenta al mismo tiempo la deriva, vea allí. 758 KFD Titrino...
  • Página 31: Parámetros Para Set

    La velocidad máxima depende de la Unidad intercambiable: Unidad intercamb. max. 5 mL 15 mL/min 10 mL 30 mL/min 20 mL 60 mL/min 50 mL 150 mL/min 758 KFD Titrino...
  • Página 32 Si está en "abs": V inicial 0.0 ml Volumen inicial absoluto (0...999.99 mL) cond. Si está en "rel": factor Factor para el cálculo del volumen inicial relativo cond. (0...±999999) Calculado según: Volumen inicial en mL = factor ∗ peso 758 KFD Titrino...
  • Página 33 El valor se utilice para correcciones de temperatura en titraciones pH. Intervalo de tiempo (1...999999 s) interv.tiempo para la adquisición de un valor medido en la lista de cond. puntos medidos. 758 KFD Titrino...
  • Página 34 Corrección de la deriva (auto, man. no) corr.deriva: auto: El valor de la deriva está tomado cond. automáticamente en el momento del comienzo. Valor de la deriva para la corrección de la deriva manual valor deriva 0.0 µl/min (0...99.9 L/min) µ cond. 758 KFD Titrino...
  • Página 35 Límite inferior para el peso (0.0...999 999) límite inf. cond. Límite superior para el peso (0.0...999 999) límite sup. 999999 cond. Salida del pulso en la línea "activate" (L6, pin 1) del puls.activación: clavijero "Remote" (prim., todo, acond., no) cond. vea página 174. 758 KFD Titrino...
  • Página 36 Se desconecta el agitador. (Desconectar agitador) Se efectúan el cálculo. Cálculos Se emiten los datos. Salida de datos Después de la titración se procede al nuevo acondicio- (Reacondicionamiento) namiento (si está seleccionado). 758 KFD Titrino...
  • Página 37 Puede obtener una buena aproximación inicial para el valor desde el punto de intersección de las tangentes: curvas "pendientes" curvas "planas" U/mV U/mV Gama regul. Gama regul. V/mL V/mL 758 KFD Titrino...
  • Página 38 El mismo tiempo de espera con una gama de incrementos de volumen extremadamente pequeños indica puntos diferentes de desconexión. Por el contrario, si se utiliza el criterio de parada "deriva", el punto de desconexión permanece el mismo. 758 KFD Titrino...
  • Página 39 El tiempo de la titración es • Ajuste la "veloc.min." más alta. demasiado largo. • Ponga la "veloc.max." más alta. • Baje la gama de regulación. Los resultados de la titración se • Ajuste la "veloc.min." más baja. dispersan demasiado. 758 KFD Titrino...
  • Página 40: Parámetros Para Meas

    (-1270...1270 mV, en pasos U(pol) 400 mV de 10 mV) Prueba del electrodo (no, sí) prueba electrodo: Prueba para electrodos polarizados. Se efectúa en el cambio del estado inactivo a una medición. "no" indica que la prueba no tiene lugar. 758 KFD Titrino...
  • Página 41 Límite inferior para el peso (0.0...999 999) límite inf. Límite superior para el peso (0.0...999 999) límite sup. 999999 Salida del pulso en la línea "activate" (L6, pin1) del puls.activación: clavijero "Remote" (sí, no) vea página 174. 758 KFD Titrino...
  • Página 42: Parámetros Para Cal

    36. El valor medido se adquiere tan pronto como se haya cumplido el primer criterio (deriva o tiempo). Si la deriva y el tiempo se han ajustado ambos en "no", se acepta inmediatamente el valor medido. Identificación del electrodo (hasta 8 caracteres ASCII) id.electrodo 758 KFD Titrino...
  • Página 43 C46: pH(as) C47: pendiente de electrodo Los datos de la calibración se pueden visualizar en cualquier momento con la tecla <CAL.DATA> y la impresión de los datos se lleva a cabo pulsando la secuencia <PRINT> <CAL.DATA> <ENTER>. 758 KFD Titrino...
  • Página 44: Parámetros Para Tip

    Para mediciones con la tecla <MEAS/HOLD>. Entrada de medida (1, 2, diff.) entrada de med: Consulta sólo con cantidades medidas pH y U. Entrada de medida 1 ó 2 o amplificador diferencial; conexión de electrodos, vea página 167. 758 KFD Titrino...
  • Página 45 "no" significa que la prueba no se realiza. Temperatura (-170.0...500.0 °C) temperatura 25.0 °C Temperatura para correcciones de temperatura en mediciones pH. Hay que entrar el valor manualmente, también si hay conectada una termosonda. 758 KFD Titrino...
  • Página 46: Cálculos Del Resultado

    RSX se encuentre en la pantalla y pulse <ENTER>. Si se memoriza una fórmula con <ENTER>, se consulta a continuación el texto, el número de decimales, la unidad y los ajustes del control de los valores límites. 758 KFD Titrino...
  • Página 47 Volumen del punto con voltaje máximo en la curva (en CAL y TIP no hay evaluación) Volumen del punto con voltaje mínimo en la curva (en CAL y TIP no hay evaluación) C70...C79 Variables transitorias para el cálculo en TIP. 758 KFD Titrino...
  • Página 48 La impresión del cálculo se puede llevar a cabo por medio de la secuencia de las teclas <PRINT><←/→> <ENTER> (pulse las teclas "Cursor" repetidamente hasta que aparezca "calc" en la pantalla). Ejemplo para un informe de cálculo: Informe de cálculo 758 KFD Titrino 10196 758.0022 usuario fecha 2001-11-21...
  • Página 49: Cálculos Estadísticos

    "borrar n": Anulación de resultados individuales con índice n. "borrar todo": Se borra la tabla completa. Índice n del resultado que va a ser borrado (1...20) borrar El resultado borrado se elimina del cálculo estadístico. 758 KFD Titrino...
  • Página 50 • Con un cambio de método, la antigua tabla de estadísticas se borra y siguen las instrucciones estadísticas para el nuevo método. • Los antiguos resultados en la tabla estadística que ya no se necesiten se pueden borrar con "borrar todo" (en la tecla <PARAM>, ">estadística", "tab.res:"). 758 KFD Titrino...
  • Página 51: Variables Comunes

    (puntos finales), CXX (variables), MNX (medias). Los valores de las variables comunes permanecen C39= para todos los métodos hasta que se sobreescriban o se borren. Se pueden visualizar los valores con la tecla <CONFIG>. Para borrar la asignación: <CLEAR>+<ENTER> 758 KFD Titrino...
  • Página 52: Emisión De Datos

    Curva volumen vs. tiempo (en SET y KFT) o curva valor medido vs. tiempo (en MEAS). cs cmpl Cálculo silo informe completo. cs breve Cálculo silo informe breve. calc Informe de cálculo con fórmulas y variables. calib Datos de la calibración. Form Feed para la impresora. 758 KFD Titrino...
  • Página 53 Los informes originales tienen guiones dobles ====, mientras que los recálculos están señalizados con guiones simples ----. La emisión de los informes se puede detener con <QUIT>. Ejemplo para una impresión: Informe completo de los resultados 758 KFD Titrino 10196 758.0022 usuario Identificación de la tarjeta insertada, vea pá-...
  • Página 54: Indicación De La Curva

    Después de la titración se puede conmutar entre la indicación de los resultados y de la curva con la tecla <CURVE>. Con las teclas <↑> y <↓> se puede trazar la curva. A la izquierda se ve el índice del punto en la primera lí- nea. Después se encuentran los valores medidos. 758 KFD Titrino...
  • Página 55: Nombre Del Usuario, Tecla

    Si no se desea que se imprima el nombre, se puede elegir „vacio“ como usuario. Borrar el nombre del usuario >borrar Introducir el nombre directamente o elegir con las nombre: teclas <←> y <→>. Con <ENTER> se borra el nombre de la lista de nombres de usuarios. 758 KFD Titrino...
  • Página 56: Tip, Proceso De Titraciones

    Confirme el bloque con <ENTER> e introduzca el 2 x <QUIT> parámetro para el bloque elegido. A continuación aparece la llamada para el segundo bloque, etc. Se pueden elegir hasta 30 bloques. Si la secuencia se ha definido totalmente, deje la consulta con <QUIT>. 758 KFD Titrino...
  • Página 57: Significado

    En TIP, el agitador no está controlada en los submétodos. Al final de TIP, el agitador se desconecta (con control de agitador en "sí"). Los parámetros de la secuencia se pueden visualizar y cambiar en cualquier momento con la tecla <PARAM>. 758 KFD Titrino...
  • Página 58 Prep Se puede preparar las buretas. Cálculos superiores en TIP. Cálculos Salida de datos Emisión de datos en TIP. TIP ya no contiene datos de determinación, es decir, las curvas se deben emitir de los submé- todos. 758 KFD Titrino...
  • Página 59 Introducción de las fórmulas vea página 42. Indicación: Es aconsejable efectuar los cálculos dentro de TIP, siempre que sea posible, porque sólo éstos se pueden volver a calcular en "seco" después de la determinación, por ejemplo, con un nuevo peso. 758 KFD Titrino...
  • Página 60: Memoria De Métodos, Teclas Y

    (hasta 8 caracteres ASCII) Si ya hay un método memorizado con la identificación dada, se le consulta si desea sobreescribir el método antiguo. Con <ENTER> se sobreescribe, con <QUIT> vuelve a la entrada de identificación del mé- todo. 758 KFD Titrino...
  • Página 61 ¡Documente sus métodos (por ejemplo con los informes "parámetros", DEF y C-FMLA)! Si tiene Ud. un PC, haga periódicamente un "backup" de sus métodos con el programa Vesuv 6.6008.XXX. ¡Si haga un "backup" en la tarjeta, presta atención a la fecha del cambio de la pila! 758 KFD Titrino...
  • Página 62: Tecla

    Los métodos se encuentran en el directorio actual de >borrar método la tarjeta. Cambiar, crear, borrar directorio >cambiar directorio >crear directorio Borrando un directorio borra todos los métodos de >borrar directorio este directorio también. (Nombre del directorio con hasta 10 caracteres) directorio: 758 KFD Titrino...
  • Página 63 • Se puede organizar los métodos en diferentes directorios; por ejemplo, se puede memorizar los métodos en directorios según el tipo de muestra o del usuario. • Se puede utilizar la tarjeta como memoria de métodos extendida. 758 KFD Titrino...
  • Página 64: Impresiones

    • Contenido del directorio actual: Impresión con las teclas <PRINT><CARD><ENTER>. • Contenido de toda la tarjeta: Impresión con las teclas <PRINT><←/→><ENTER> pulse las teclas cursor repetidamente hasta que aparezca "tarjeta" en la pantalla. Referencias de pedido Tarjeta de memoria con 128 KByte..............6.2245.010 758 KFD Titrino...
  • Página 65: Datos De Calibración, Tecla

    CAL, esta entrada se acepta automáticamente después de la calibración. La impresión de los datos de calibración para la entrada de medida actual se puede efectuar en cualquier momento con la secuencia de las teclas <PRINT><CAL.DATA><ENTER>. 758 KFD Titrino...
  • Página 66: Datos De La Muestra Actual, Tecla

    Entrada por medio del teclado, de una balanza o con un lector de código de barras. Unidad del peso (g, mg, mL, L, pc, ninguna o hasta 5 µ unidad del peso: caracteres ASCII) Selección con <←/→>. 758 KFD Titrino...
  • Página 67: Memoria Silo Para Datos De Muestras

    En operaciones mixtas, entrada manual de identificaciones y pesos por medio de una balanza, se envían los datos desde la balanza a la línea en la que se está editando. Debe ser confirmado con <ENTER> en el Titrino. 758 KFD Titrino...
  • Página 68 El acceso está bloqueado durante el proceso. Como indicación de que una línea ha sido borrada aparecen las consultas con "∗". El signo ∗ indica que la línea ha sido borrada. Las líneas borradas se pueden reactivar cuando se vuelven a editar. 758 KFD Titrino...
  • Página 69 Memoria silo conteniendo 35 líneas. 6 de las 35 líneas han sido La siguiente línea libre es la 36. procesadas. Las líneas procesadas han sido copiadas al final de la memoria silo: el silo está lleno hasta la línea 41. 758 KFD Titrino...
  • Página 70: Memorización De Resultados Y Cálculos Silo

    ). Se memorizan el nombre del resultado, valor y unidad. El lugar que un valor necesita se puede calcular con la ayuda de los siguientes datos: Resultado con su texto (8 caracteres) y unidad (5 caracteres): 32 bytes Valor medido C40, valor sin unidad: 22 bytes 758 KFD Titrino...
  • Página 71: Cálculos Silo

    2.17 Memorización de resultados y cálculos silo Después de trabajar con algunas muestras, el informe silo puede tener el siguiente aspecto (impresión con <PRINT><SILO><ENTER>): 758 KFD Titrino 10196 758.0022 fecha 2001-11-21 hora 15:03 >silo ciclo datos: memorizar resultados: sí sl método id 1/C21 id 2/C22 id 3/C23 Líneas silo...
  • Página 72 2.17 Memorización de resultados y cálculos silo Partiendo del siguiente informe silo: 758 KFD Titrino 10196 758.0022 fecha 2001-11-21 hora 15:03 >silo ciclo datos: memorizar resultados: sí método id 1/C21 id 2/C22 id 3/C23 C00 11-2 A/12 99-08-12 0.903g 2.6427ml 1.48%...
  • Página 73 3. Cálculo de la media (MNX) 4. Asignaciones de los resultados silo C24 y C25 5. Cálculos silo 6. Asignaciones de las medias de los cálculos silo en C26 y C27 7. Asignaciones de las variables comunes 758 KFD Titrino...
  • Página 74: Dosificación Manual Y Preparación De Las Buretas

    • volumen del tubo de aspiración • volumen del cilindro • volumen del tubo de dosificación Se puede expeler el volumen del cilindro o en la punta o en la botella. Los parámetros se introducen con <CONFIG>, vea página 13. 758 KFD Titrino...
  • Página 75: Operation Via Rs232 Interface (Páginas Verdes, En Inglés)

    Entries which start with $, trigger some- thing. They are thus called triggers. Values of objects can not only be queried, they can also be modified. Values are always entered in quotes, for example: &Config.Aux.Language "english" 758 KFD Titrino...
  • Página 76: Call Up Of Objects

    'Language' at this level: ..L quire n+1 preceding dots. Change from node 'Language' via the root to If you must jump back to the root, enter a pre- node 'Mode': &M ceding &. 758 KFD Titrino...
  • Página 77: Triggers

    Examples: Querying the value of the baud rate: &Config.RSSet.Baud $Q Querying all values of the node "RSSet": &Config.RSSet $Q Querying the path of the node "RSSet": &Config.RSSet $Q.P Start mode: &Mode $G Querying the detailed status: $D 758 KFD Titrino...
  • Página 78: Status Messages

    Preparing buret X, X=0, 1, 2. $G .Prep.X.Active: Emptying buret X, X=1, 2. .Empty.X.Active: In TIP, its global status as well as the step number (X) is available. Instrument at the beginning or at the end of a TIP. $G .TIP.X .Inac: 758 KFD Titrino...
  • Página 79 $S .Mode.XXXX.QuickMeas: The instrument gives the status from which it has been stopped. The detailed status information is therefore identical the information for the global status Violation of monitored limits with action "end" give the status message $S.Mode.XXX.Inac;EYYY. 758 KFD Titrino...
  • Página 80: Error Messages

    Exit: Send new command. Value has been corrected automatically. Exit: Send new command. Instrument at the end of the titration and sample data is edited; the instrument at rest or editing during filling. Exit: &m $S. 758 KFD Titrino...
  • Página 81 Exit: Send a silo entry. E133 Silo full. Exit: Send new command. E134 No method. A method, which is required from the silo memory or in TIP, does not exist. Exit: The error message disappears on next startup. 758 KFD Titrino...
  • Página 82 Exit: Send new command. E183 A directory with the same name exists already on the memory card. Exit: Send new command. E196 Result is out of limits. Exit: The error message disappears on next start or on recalculation. 758 KFD Titrino...
  • Página 83 <CLEAR>. The dosing element will be initialized. E282 Overload in dosing element: cock is blocked. Exit: <STOP>, <QUIT>. Rectify fault at the wet part. Mount it again onto the dos driver and press <CLEAR>. The dosing element will be initialized. 758 KFD Titrino...
  • Página 84: Remote Control Commands

    Administration of the memory card ├ Config Instrument configuration ├ SmplData Sample specific data ├ Hotkey Keys with direct access ├ Info Current Data ├ Assembly Component data ├ Setup Setting the operating mode └ Diagnosis Diagnostics program 758 KFD Titrino...
  • Página 85 │ │ ├ .C70 for C70 RSX,EPX,CXX 3.2.2.10. │ │ ├ up to C79 ├ .CFmla Calculation constants │ ├ Calculation constant C01 │ │ ├ .Value Input of value 0...±999999 3.2.2.11. │ ├ up to C19 758 KFD Titrino...
  • Página 86 Status of limit control ON, OFF ditto │ │ ├ .LoLim Lower limit 0.0...999 999 ditto │ │ ├ .UpLim Upper limit 0.0...999 999 ditto │ ├ .ActPulse Output of a pulse first, all, cond., OFF 3.2.2.30. 758 KFD Titrino...
  • Página 87 .LoLim Lower limit 0.0...999 999 ditto │ │ ├ .UpLim Upper limit 0.0...999 999 ditto │ ├ .Oven KF Oven connected COM1, COM2, no 3.2.2.33. │ ├ .ActPulse Output of a pulse first, all, cond., OFF 3.2.2.30. 758 KFD Titrino...
  • Página 88 Status of statistics calculation ON, OFF 3.2.2.26. │ ├ .MeanN No.of individual determinations 2...20 ditto │ ├ .ResTab Result table │ │ ├ .Select original,delete n,delete all ditto │ │ ├ .DelN Deletion of individual results 1...20 ditto 758 KFD Titrino...
  • Página 89 Measuring input 1, 2, diff. ditto │ ├ .Ipol Polarization current 0...1...±127 ditto │ ├ .Upol Polarization voltage 0...400...±1270 ditto │ ├ .PolElectrTest Test for polarized electrodes ON, OFF ditto │ ├ .Temp Titration temperature –170.0...25.0...500.0 ditto 758 KFD Titrino...
  • Página 90 │ │ ├ .DosUnit Dosing element read only ditto │ │ ├ .Bytes Method size in bytes read only ditto │ │ ├ .Checksum Checksum of method read only ditto ├ for each method 758 KFD Titrino...
  • Página 91: Memory Card

    │ │ │ ├ .DosUnit Dosing element read only ditto │ │ │ ├ .Bytes Method size in bytes read only ditto │ │ │ ├ .Checksum Checksum of method read only ditto │ │ ├ for each method 758 KFD Titrino...
  • Página 92 Time interval for validation 1...7...9999 ditto │ │ ├ .Counter Time counter 0...9999 ditto │ ├ .Service Monitoring of Metrohm service 3.2.2.64. │ │ ├ .Status Status of service monitoring ON, OFF ditto │ │ ├ .Date Date of next service...
  • Página 93 │ │ │ │ ├ .Diam Diameter 0.1...2.0...9.9 ditto │ │ │ ├ .DosRate Dosing rate 0.01...150, max. 3.2.2.91. │ │ │ ├ .FillRate Filling rate 0.01...150, max. ditto │ ├ Buret D2 as for D1 758 KFD Titrino...
  • Página 94 .DelLine Delete silo line 3.2.2.96. │ │ ├ .LineNum Line number 1...255, OFF ditto │ ├ .DelAll Delete silo line 3.2.2.97. │ ├ .CycleLines Cycle lines ON, OFF 3.2.2.98. │ ├ .SaveLines Save results ON, OFF 3.2.2.99. 758 KFD Titrino...
  • Página 95 10 ASCII char ditto │ ├ .DelAll Delete all users ditto │ ├ .List List of users │ │ ├ User 1 │ │ │ ├ .Name Name of user read only ditto │ │ ├ up to 99 758 KFD Titrino...
  • Página 96 │ │ │ for each measuring point │ ├ .TitrResults Titration results │ ├ Calculated results │ │ ├ result │ │ │ ├ .Value Value read only 3.2.2.106. │ │ ├ up to 9 results 758 KFD Titrino...
  • Página 97 │ │ ├ .Mean Mean value read only ditto │ │ ├ .Std Absolute standard deviation read only ditto │ │ ├ .RelStd Relative standard deviation read only ditto │ ├ .C27 as for C26 758 KFD Titrino...
  • Página 98 │ ├ .Comport Comport │ │ ├ .Number COM where PC is connected read only 3.2.2.117. │ ├ .Assembly Assembly │ ├ CycleTime Cycle time read only 3.2.2.118. │ ├ Volume of Exchange/Dosing unit read only ditto 758 KFD Titrino...
  • Página 99 3.2.2.126. │ ├ .SetLines Set I/O lines ditto │ │ ├ Signal on L0 active,inactive,pulse,OFF ditto │ │ ├ up to L13 │ ├ .ResetLines Reset I/O lines ditto │ ├ .Stirrer Stirrer ON, OFF 3.2.2.127. │ 758 KFD Titrino...
  • Página 100 │ │ ├ .dMeasdt Measured value drift ON, OFF ditto │ │ ├ .dMeasdV 1st deviation of titration curve ON, OFF ditto │ │ ├ .ERC without meaning │ │ ├ Temp. as secondary meas. ON, OFF ditto │ 758 KFD Titrino...
  • Página 101 .Initialise Set default values 3.2.2.141. │ ├ .Select Selection of branch ActMeth,Config,Silo,Calib │ Assembly,Setup,All ditto ├ .RamInit Initialization of working mem. 3.2.2.142. ├ .InstrNo Device Identification 3.2.2.143. ├ .Value Input of device identification 8 ASCII characters ditto 758 KFD Titrino...
  • Página 102 3.2 Remote control commands &Diagnose Object Description Input range Reference & Root ├ Diagnose Diagnose ├ .Report Output of adjustment parameters 3.2.2.144. 758 KFD Titrino...
  • Página 103: Description Of The Remote Control Commands

    Entry of formulas. Rules for formula entry, see page 42. Example: "(EP2–EP1)∗C01/C00" In addition to the formula, a text for result output, the number of decimal places and a unit for the result output can be selected. "No unit" is selected with the blank string. 758 KFD Titrino...
  • Página 104 In addition, the statistics calculation must be swit- ched on, see 3.2.2.26. Rules for statistics calculations see page 46. 3.2.2.10. Mode.Def.TempVar.C70 RSX, EPX, CXX etc. up to .C79 Assignment of temporary variables in a submethod for calculations in TIP. 758 KFD Titrino...
  • Página 105 "auto" means the titration direction is determined automatically by the instru- ment. If 2 EP's have been set in a SET titration, the titration direction is given by the two EP's. The entry of the titration direction is then invalid. 758 KFD Titrino...
  • Página 106 With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in steps of 10 mV. Besides .PolElectrTest is valid. If the test for polarized electrodes is switched on, it is performed on change- over from the inactive state to an active state (titration or conditioning). 758 KFD Titrino...
  • Página 107 All results of the corresponding line in the statistics table are deleted. Specification of the line number in .ResTab.DelN. delete all: Clear entire statistics table. The results can not be reac- tivated. .ResTab.DelN: Specification of the line number to be deleted. 758 KFD Titrino...
  • Página 108 Parameters for KFT: Setting of the EP, resp. Control point as a pH value, in mV (with U and Ipol) or µA (with Upol). The corresponding unit can be read with .UnitEP. With KFT the standard value for Ipol is 250 mV, for Upol 25 µA. 758 KFD Titrino...
  • Página 109 With Upol the inquiry for the polarization voltage in mV (Upol) is valid. Entry in steps on 10 mV. Besides .PolElectrTest is valid. If the test for polarized electrodes is switched on, it is performed on change- over from the inactive state to the measurement. 758 KFD Titrino...
  • Página 110 Parameters for CAL: Electrode identification. It is classified under calibration data, see 3.2.2.102. 3.2.2.43. Mode.Parameter.Calibration.SmplChanger ON, OFF Parameters for CAL: Calibration at Titrino. With "ON", there are no hold points in the calibration sequence for entries, the first buffer is measured directly. 758 KFD Titrino...
  • Página 111 (Upol) applies. Entry in steps of 10 mV. .PolElectrTest also ap- plies. If the test for polarized electrodes is switched on (ON), it will be per- formed on the change from the inactive state to an active state. The tempera- ture applies to pH measurements. 758 KFD Titrino...
  • Página 112 Administration of the method memory of the current directory of the memory card: load, store and delete methods. The action is carried out, if "$G" is transmitted to the corresponding node. Do not use blank characters before and after method name! 758 KFD Titrino...
  • Página 113 Do not use blank characters before and after name! 3.2.2.58. MemoryCard.FreeMem read only Free memory on the card. $Q sends number of free bytes, e.g. "4928". 3.2.2.59. MemoryCard.BatteryChange MemoryCard.BatteryChange.Date YYYY-MM-DD Date for changing battery. The date will be set with "$G". 758 KFD Titrino...
  • Página 114 Time counter in days since last calibration. The counter is reset to zero if a new calibration is carried out or if the calibration for the corresponding measuring input is entered manually. 3.2.2.64. Config.Monitoring.Service.Status ON, OFF Config.Monitoring.Service.Date YYYY-MM-DD Monitoring of service interval. 758 KFD Titrino...
  • Página 115 .Barcode: Select target in Titrino where you wish to have the string from the barcode reader. "input" means that the string comes into the field where the cursor is currently placed. 3.2.2.71. Config.Aux.Language english, deutsch, francais, español, italiano, portugese, svenska Selection of the dialog language. 758 KFD Titrino...
  • Página 116 (3.2.2.138) is used at the same time. If a name has been entered, it will be printed out in the result report (full, short). 3.2.2.78. Config.Aux.Prog read only Output of the program version. The Titrino sends "758.0022" on requests with $Q. 758 KFD Titrino...
  • Página 117 Time interval in min for automatic warning that a preparation has to be exe- cuted. 3.2.2.85. Config.DosPrep.DX(.Dosimat).V 0...3.5...99999.99 Volume in mL, which will be expelled at the preparation. Start sequence see 3.2.2.121. 3.2.2.86. Config.DosPrep.DX(.Dosimat).Repeat 1...2...9 Number of cycles for expelling the volume at the preparation. Start sequence see 3.2.2.121. 758 KFD Titrino...
  • Página 118 5 ASCII characters Current sample data. The identifications Id1...Id3 can be used in formulas as sample-specific calcu- lation constants C21...C23. If "no unit" is desired for the unit of the sample size, the blank string must be entered. 758 KFD Titrino...
  • Página 119 Deletes the entire silo memory. Must be triggered with $G. 3.2.2.98. SmplData.ONSilo.CycleLines ON, OFF Silo data cycling. With "ON", executed lines are copied to the next free silo lines, see page 65. Exercise caution if you edit the silo memory during the determinations! 758 KFD Titrino...
  • Página 120 Short report of the silo calculations. prep: Preparation report. calc: Calculation report of the current method. act dir: Methods of the current directory of the memory card. mem card: All methods of the memory card. all: All reports. Form feed on printer. 758 KFD Titrino...
  • Página 121 Info.DetermData.ExV read only/read+write Info.DetermData.MPList.1.X read only/read+write Info.DetermData.MPList.1.Y read only/read+write Info.DetermData.MPList.1.Z1 read only/read+write Info.DetermData.MPList.1.Z2 read only/read+write for every measuring point Determination data in hexadecimal format. A measuring point list is available in mode SET, KFT and MEAS. 758 KFD Titrino...
  • Página 122 C48: Voltage at maximum voltage (not valid in CAL and TIP) C49: Voltage at minimum voltage (not valid in CAL and TIP) DTime: Time for the drift correction in SET/KFT with conditioning. .TempVar: Temporary variables in TIP corresponding to the assignments in the submethods. 758 KFD Titrino...
  • Página 123 C26.ActN: Number of single results "3" C26.Mean: Mean (decimal places as for the result itself) "3.421" C26.Std: Standard deviation (decimal places as for the result + 1) "0.0231" C26.RelStd: Relative standard deviation (in %, 2 decimal places) "0.14" 758 KFD Titrino...
  • Página 124 QuickMeas. It is incremented as long as the instrument remains switched on. 3.2.2.111. Info.ActualInfo.Assembly.Counter.V read only Info.ActualInfo.Assembly.Counter.Clear $Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the volume counter is set to zero. 3.2.2.112. Info.ActualInfo.Assembly.Meas read only $Q sends the current measured value from the assembly. 758 KFD Titrino...
  • Página 125 .MeasPt.Y"3.654" Volume (SET, KFT) of the MPList in mL .MeasPt.Z1"6.34" Measured value (SET, KFT, MEAS) of the MPList, format depends on the measured quantity .MeasPt.Z2"25.8" Temperature of the MPList in °C .EP.X"1.234" Volume coordinate of the EP .EP.Y"5.34" Measured value coordinate of the EP 758 KFD Titrino...
  • Página 126 D0, external D1, external D2 Selection of buret for assembly functions. 3.2.2.120. Assembly.Bur.Empty $G, $S, $H, $C Starts the function "empty". Only possible with the burette type "Dosino". The parameters are under the function &Config.DosPrep, see 3.2.2.89 up to 3.2.2.91. 758 KFD Titrino...
  • Página 127 Control. When the measuring function is switched on, no method can be started at the Titrino. .Input: Selection of the potentiometric measuring input 1, 2, diff., polar- ized electrodes or temperature. .Ipol: Polarization current in µA. .Upol: Polarization potential in mV, entry in steps of 10 mV. 758 KFD Titrino...
  • Página 128 L8–13 .ResetLines: Lines are set to the inactive status (= high). 3.2.2.127. Assembly.Stirrer ON, OFF Switching stirrer ON/OFF. 3.2.2.128. Setup.Comport 1, 2, 1&2 Selects the Titrino COM for the output of automatic info: &Setup.Keycode &Setup.Trace &Setup.SendMeas &Setup.AutoInfo 758 KFD Titrino...
  • Página 129 .UserMeth.Store: Disable "store" in <user meth> key .UserMeth.Delete: Disable "delete" in <user meth> key .Display: Disable the display, i.e. it will not be written to by the device pro- gram of the Titrino and can be operated from the computer. 758 KFD Titrino...
  • Página 130 ON, OFF Setup.SendMeas.Titrator.Meas ON, OFF Setup.SendMeas.Titrator.dVdt ON, OFF Setup.SendMeas.Titrator.dMeasdt ON, OFF Setup.SendMeas.Titrator.dMeasdV ON, OFF Setup.SendMeas.Titrator.T ON, OFF Selection of the values from the titrator which are sent in the set time interval (see 3.2.2.134, formats see 3.2.2.113): 758 KFD Titrino...
  • Página 131 Final: End of determination (or submethod), the final steps will be carried out. .T.E Error. Message together with error number, see page 76ff. .T.H Hold: Status 'Hold' has been reached. .T.C Continue: Continue after hold. .T.O Conditioning OK: EP reached (in SET, KFT with conditioning). 758 KFD Titrino...
  • Página 132 With auto, the smallest measured value is taken and the next smaller tick defines the beginning of the scale; the next greater tick to the greatest measured value is the end of the scale. 758 KFD Titrino...
  • Página 133 Initializes instrument, see page 153. All parameters are set to their default value and error messages are cleared. The user and silo memories will be de- leted. The user memory contains the default user methods from Metrohm. 3.2.2.143. Setup.InstrNo Setup.InstrNo.Value serial number, 8 ASCII characters Instrument identification for report output.
  • Página 134: Properties Of The Rs 232 Interface

    Parity Bit 1 or 2 Stop Bit Only a shielded data cable (for example, METROHM D.104.0201) may be used to couple the Titrino with foreign devices. The cable shield must be properly grounded on both instruments (pay attention to current loops; always ground in a star-head formation).
  • Página 135 Handshake output ports (DTR, RTS) are set by the Titrino. The Titrino has an input buffer which can accept up to 512 characters. Titrino as Receiver : Titrino external device max. 128 characters Data output 384 characters XOFF Data input Time 758 KFD Titrino...
  • Página 136 6 s by XOFF, E43 appears in the display. Hardware-Handshake, HWs Titrino as Receiver : Titrino external device Time Titrino as Sender: Titrino external device Time The data flow can be interrupted by deactivating the CTS line. 758 KFD Titrino...
  • Página 137: Pin Assignment

    (<-3 V): OFF state voltage positive (>+3 V): ON state In the transitional range from +3 V to -3 V the signal state is undefined. Driver 14C88 according to EIA RS 232C specification Receiver 14C89 " " 758 KFD Titrino...
  • Página 138 Contact arrangement at plug (female) for RS 232C socket (male) View of soldered side of plug Ordering numbers: K.210.0381 and K.210.9045 No liability whatsoever will be accepted for damage or injury caused by im- proper interconnection of instruments. 758 KFD Titrino...
  • Página 139: Qué Hacer Si La Transmisión De Datos No Funciona

    • ¿Está su cable correctamente cableado? (El DTR de la impresora ha de ser conectado al CTS del Titrino). • Ajuste el handshake a "HWs" en el Titrino. Configure la impresora de modo que esté ajustado el DTR (normalmente con interruptor DIP). 758 KFD Titrino...
  • Página 141: Mensajes De Errores, Corrección De Averías

    Salida: <QUIT> y asegúrese de que los parámetros corres- pondientes son los mismos en ambos aparatos. Overrun error. Al menos 1 carácter no se ha podido leer. error 38 Salida: <QUIT> El tampón receptor del Titrino está completo (>128 caracte- error 39 res). Salida: <QUIT> 758 KFD Titrino...
  • Página 142 Salida: <CLEAR>. Atención: Los datos de esta línea silo no están memorizados En la secuencia de la calibración la diferencia de voltaje entre mismo tampón el primer tampón y el segundo es <6 mV. Salida: <QUIT> y cambio de tampón o <STOP> (parada de la calibración). 758 KFD Titrino...
  • Página 143 (el intervalo de aviso está expirado). Salida: Comienzo de la preparación con <START> o fingir el aviso con <CLEAR>. La preparación o el vacío de la bureta de titración fue parado prep: parada manual con <STOP>. 758 KFD Titrino...
  • Página 144 . La caja Remote se ha conectado con el Titrino en marcha. . Defecto en el Titrino. . Defecto en la caja Remote. Corrección: Ajuste "Remote-Box:no" en <CONFIG>, >apara- tos periféricos, desconecte el Titrino, quite la caja Remote y conecte el Titrino. Llame al servicio Metrohm. 758 KFD Titrino...
  • Página 145 El voltaje de la pila está en la zona de 2.37...2.64 V. volt.pila de tarj.bajo Corrección: <CLEAR> y cambie pila el más rápido que posible. La titración se detiene porque se ha alcanzado el volumen de V parada alcanzado parada. 758 KFD Titrino...
  • Página 146: Diagnóstico

    4.2 Diagnóstico 4.2.1 Introducción El 758 KFD Titrino es un aparato de medición de una gran precisión y digno de confian- za. Gracias a su fuerte construcción es prácticamente imposible que sus funciones se vean alteradas por influencias externas, mecánicas o eléctricas.
  • Página 147: Aparatos Necesarios

    • La secuencia de la prueba se puede detener en cualquier momento pulsando la tecla <5> y después continuarla. • El bloque 2 se quita con la tecla <CLEAR>. d i a g n o s e p r e s s k e y 0 . . . 9 758 KFD Titrino...
  • Página 148: Prueba Del Teclado

    Placa frontal 758 Fig. 2 Teclado 758 • En bloque 1 so deja pulsando <CLEAR> dos veces. d i a g n o s e p r e s s k e y 0 . . . 9 758 KFD Titrino...
  • Página 149: Código Del Cilindro, Fecha Y Hora

    - t i m e r t e s t o . k . • Pulse <CLEAR>. d i a g n o s e p r e s s k e y 0 . . . 9 758 KFD Titrino...
  • Página 150: Prueba De Las Entradas De Alto Ohmiaje

    6.1 Prueba de las entradas de alto ohmiaje • Conecte la entrada de medición ‘Ind I’ a un calibrador de tensiones (p. ej. simula- dor pH Metrohm 773), utilizando un cable de alto ohmiaje (p. ej. 6.2108.060). Pon- ga el calibrador a 0 mV.
  • Página 151: Prueba Del Polarizador

    . k . • Pulse <CLEAR> d i a g n o s e p r e s s k e y 0 . . . 9 • Quite 767 o el cable y la década de resistencia. 758 KFD Titrino...
  • Página 152: Entradas Y Salidas Externas

    Entradas y salidas externas Esta prueba es necesaria sólo si el 758 KFD Titrino se utiliza conectado a otros aparatos por medio del enchufe a la conexión ‘Remote‘ o si se emplea la caja Remote 6.2148.000 para la conexión de un teclado del PC o un lector de código de barras.
  • Página 153 Por razones de integridad se describe aquí el procedimiento total. Si no desea una prueba de diagnóstico de RS 232 continúe con el punto 9. Fig. 5 Conexiones en la ficha de prueba 3.496.8560 • Pulse <5>. RS232 test 1...2 758 KFD Titrino...
  • Página 154 . k . • Quita la ficha de prueba. • Pulse <CLEAR>. RS232 test 1...2 • Pulse <CLEAR>. d i a g n o s e p r e s s k e y 0 . . . 9 758 KFD Titrino...
  • Página 155: Prueba De La Tarjeta De Memoria

    Prueba de la tarjeta de memoria • Pulse <9>. memory card test • Introduzca la tarjeta de memoria. Se necesita una tarjeta formatada por el 758 KFD Titrino. Si está formatada por otro Titrino, aparece el mensaje „memory card 000 KB o.k.“. • Pulse <ENTER>.
  • Página 156 • Pulse <DOS> (del aparato) al mismo tiempo y utilice un cronómetro para medir el tiempo en que se expulsa 1/10 del volumen del cilindro. El tiempo debería ser 90...110 s, aproximadamente. • Ponga el potenciómetro 'dV/dt' al tope derecho. • Pulse <FILL>. 758 KFD Titrino...
  • Página 157: Vuelta A La Posición Inicial

    4.2 Diagnóstico 11 Bureta externa D 1 o D 2, resp. Esta prueba es necesario sólo si el 758 KFD Titrino se utiliza conectado con buretas ex- ternas (685 Bureta o 700 Dosino). Con la ayuda de la función <PREP> (ver en la pagina 70) se puede probar el acciona- miento del eje y el cambio de la posición del grifo de la bureta interna.
  • Página 158: Inicialización Y Prueba Ram

    <CLEAR>. Los valores de inicialización se cargan automáticamente. De este modo el aparato puede medir. Hay que contar, sin embargo, con una posible pérdida de exactitud. El servicio de Metrohm podría realizar un nuevo y óptimo reajuste. El mensaje de error aparece siempre después de conectar el aparato...
  • Página 159: Desbloqueo De Un Eje Bloqueado Con Unidad Intercambiable

    ¡Los circuitos electrónicos quedan ahora accesibles! ¡Por ningún motivo debe tocarlos! • Quite el eje del tope mecánico girando la rueda dentada grande. (En caso de que el motor esté inactivo, ponga el eje a mano en la posición cero.) 758 KFD Titrino...
  • Página 160: Preparaciones

    5.1.1 Titrino con agitador o Ti Stand Cable del agitador Agitador o Ti Stand Teclado 6.2130.080 También se pueden conectar la Varilla agitadora 802 o el Ti Stand 727 o 703 con cable 6.2108.100 en lugar del Agitador 728. 758 KFD Titrino...
  • Página 161: Titrino Con Buretas Externas

    5.1 Instalación y conexiones de los instrumentos 5.1.2 Titrino con buretas externas 685 Dosimat Cable 6.2134.030 700 Dosino El Dosino 2.700.0020 se conecta directamente, mientras el Dosino 2.700.0010 necesita el cable 6.2134.020. 758 KFD Titrino...
  • Página 162: Conexión De Una Impresora

    DPU-414 Data bit: ninguno Stop bit: Paridad: ninguna Handshake: Transm.a: Seiko Custom 6.2134.110 Baud Rate: 9600 ninguna, la versión Metrohm esta DP40-S4N Data Bit: configurado correctamente Stop Bit: IDP-560 EMULATION FONT MAP =GERMANY Paridad: ninguna PRINT=REVERSE Handshake: LITTLE CR CODE=VOID Transm.a:...
  • Página 163: Conexión De Una Balanza

    Con la ayuda de una unidad especial de entradas - proporcionada por el fabricante de la balanza - se pueden introducir desde la balanza y junto con el peso, identificaciones de las muestras y 758 KFD Titrino...
  • Página 164: Conexión De Un Horno Kf

    La conexión de los dos interfaces RS (cable 6.2125.110) sólo es ne- cesaria si desea obtener los resultados del horno en la impresión del Titrino 758. ¡Del horno no se puede obtener ningún resultado! Cable Remote 6.2141.010 Cable RS 232 6.2125.110 758 KFD Titrino...
  • Página 165: Conexión De Un Cambiador De Muestras

    "cambiador muestras" a "sí". • En las conexiones con el cambiador de muestras se debe ajustar "comienzo auto" a "no" bajo la tecla <CONFIG>. La orden del comienzo se da desde el Cambiador de muestras. 758 KFD Titrino...
  • Página 166: Conexión Del Oven Sample Processor 774

    Además se puede cargar un método determinado a través de la co- nexión RS en el Titrino desde el Oven Sample Processor durante el desarrollo de una secuencia. Por medio de la conexión RS se puede consultar también los parámetros del horno desde el Titrino. 758 KFD Titrino...
  • Página 167: Conexión De Un Ordenador

    DIN de 5 tomas. Es indispensable que el lector de código de barras emule a un teclado de ordenador. Al conectar al mismo tiempo un lector y un teclado, el lector ha de disponer de enchufe de conexión T. El teclado de ordenador quedará entonces comuni- cado a esta conexión del lector. 758 KFD Titrino...
  • Página 168: Observaciones

    Elija el teclado específico de su país. Si el Titrino no reconoce su teclado, elija otro teclado similar (compruebe p.ej. la 2ª distri- bución de las teclas numéricas). Los signos específicos del país posiblemente no estén correctamente adaptados. 758 KFD Titrino...
  • Página 169: Manejo Por Medio Del Teclado Del Ordenador:

    <PARAM> <PREP> Alt E <PRINT> Alt P Elija impresión con → ← <QUIT> <REPORTS> Alt O Impresión de informe: Alt P + Alt O <SILO> Sí/no <SMPL DATA> <START> <STATISTICS> Sí/no <STOP> <USER METH> <USER> Alt U 758 KFD Titrino...
  • Página 170 Overlay. Vd. puede copiar esta representación, cortar la parte central y colocarla sobre las teclas de su ordena- dor. Silo Clear Statistics Start Stop Meas/Hold Config Param SmplData 758 KFD Titrino...
  • Página 171: Conexión De Electrodos, Instalación Del Recipiente De Titración

    Con el Cambiador de medición muestras Ind I Electrodo pH 6.0133.100 Electrodo pH 6.0130.100 Ind II Electrodo de referencia con Electrodo de referencia con doble blindaje Ag/AgCl doble blindaje Ag/AgCl 6.0729.100 6.0729.110 Electrodo auxiliar 6.0301.100 Electrodo auxiliar 6.0302.110 758 KFD Titrino...
  • Página 172 CO pueden alterar la titración. • una grande distancia de la adición del reactivo al electrodo, vea el dibujo abajo. Agitador Electrodo Bureta Preste atención a la direción de rotación del agitador. 758 KFD Titrino...
  • Página 173: Instalación Del Recipiente De Titración

    6.0338.100 Tapón Boquilla roscada 6.2730.030 6.2730.010 Tapón para septo 18 mm 6.2730.020 Septo 6.1448.010 Juego de juntas 6.1244.040 Recipiente de titración, parte superior 6.1414.030 Recipiente de titración, parte inferior Varilla de agitación 20...90 mL magnética 6.1415.220 6.1903.020 758 KFD Titrino...
  • Página 174: Apéndice

    ≤ 0.3 % en relación al volumen nominal Buretas de titración 2 buretas externas y 1 bureta interna externa: 685 Dosimat o 700 Dosino Buretas auxiliares 2 buretas adicionales: 776 o 765 Dosimat Material Caja Polibutilenotereftalato (PBTP) Cubierto del teclado Policarbonato (PC) 758 KFD Titrino...
  • Página 175: Especificaciones De Seguridad

    Flash legible Memoria interna Memoria para hasta 100 métodos Banco de datos con 25 métodos Metrohm Memoria silo para datos de muestra y para resultados Interfaces RS232 2 interfaces separados para la conexión de una impresora y de una balanza o para la conexión de un ordenador: El Ti- trino es perfectamente controlable por un aparato externo Entradas/Salidas analógicas "Remote"...
  • Página 176 2 x 1 ATH (sólo se puede remplazar por el servicio Me- trohm por el mismo tipo) Protección adicional electrónica contra sobrecarga. Dimensiones con la unidad intercambiable Anchura 150 mm Altura 450 mm Profundidad 275 mm Peso, incluido teclado aproximadamente 3.6 kg 758 KFD Titrino...
  • Página 177: Asignación De Las Tomas De La Ficha "Remote

    No se utilizan pin 24 (Input 6) +5 V pin 12 (Input 7) Salidas pin 5 (Output 0) Ready inactive pin 18 (Output 1) Conditioning ok, activo cuando acondicionamiento ok pin 4 (Output 2) Titración, activo durante la titración 758 KFD Titrino...
  • Página 178 I ≤ 200 mA +5 V pin 14 0 V: activo 5 V: inactivo pin 25 Número de pedidos: K.210.9004 (vaina) y K.210.002 No nos hacemos responsables si el aparato resulta dañado debido a una incorrecta interconexión con otros instrumentos. 758 KFD Titrino...
  • Página 179: Estado De Las Líneas De La Ficha "Remote

    6.2 Asignación de las tomas de la ficha "Remote" 6.2.1 Estado de las líneas de la ficha "Remote" 758 KFD Titrino...
  • Página 180: Configuraciones Posibles Del "Puls.activación" En Modos Set, Kft Y Cal

    6.2 Asignación de las tomas de la ficha "Remote" 6.2.2 Configuraciones posibles del "puls.activación" en modos SET, KFT y CAL 758 KFD Titrino...
  • Página 181: Métodos De Usuario

    <DEF>, >impresión. Si necesita los resultados en otra unidad, hay que adaptar eventualmente los valores de cálculo con la tecla <C-FMLA>. Los métodos siguientes están disponibles: 758 KFD Titrino 10196 758.0022 usuario...
  • Página 182: Título Kf Con Agua O Metanol Estándar "H2Otiter

    Karl-Fischer-Titration according to corr.deriva: Eugen Scholz, Riedel de Haën, Seelze, Ger- llamada ident: llamada peso: many límites peso: METROHM Application Bulletin No. 77: horno: Karl Fischer Water Determinations puls.activación: ------------ 758 KFD Titrino 758.0022 fecha 2001-11-21 hora...
  • Página 183: Título Kf Con Tartrato Sódico "Tartiter

    HYDRANAL® Practical Course, Water rea- corr.deriva: gents for Karl-Fischer-Titration according to llamada ident: Eugen Scholz, Riedel de Haën, Seelze, Ger- llamada peso: many límites peso: METROHM Application Bulletin No. 77: horno: puls.activación: Karl Fischer Water Determinations ------------ 758 KFD Titrino 758.0022 fecha 2001-11-21...
  • Página 184: Valor En Blanco Para Kft "Blank_Kf

    HYDRANAL® Practical Course, Water re- llamada peso: agents for Karl-Fischer-Titration according to límites peso: Eugen Scholz, Riedel de Haën, Seelze, Ger- horno: many puls.activación: METROHM Application Bulletin No. 77: ------------ Karl Fischer Water Determinations 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 18:01...
  • Página 185: Titración Kf Con Deducción De Un Valor En Blanco

    Karl-Fischer-Titration according to llamada peso: todo Eugen Scholz, Riedel de Haën, Seelze, Ger- límites peso: many horno: METROHM Application Bulletin No. 77: puls.activación: Karl Fischer Water Determinations ------------ 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 18:04...
  • Página 186 6.3 Métodos de usuario 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 18:04 Ipol KF-Blank >fórmula Water=(EP1-C38)*C39*C01/C00/C02 RS1 texto Water RS1 num.decimales RS1 unidad: RS1 control val.límite: Titer=C39 RS2 texto Titer RS2 num.decimales RS2 unidad: mg/ml RS2 control val.límite: Blank=C38 RS3 texto Blank RS3 num.decimales...
  • Página 187: Titración Kf Sin Deducción De Un Valor En Blanco "Kf

    Karl-Fischer-Titration according to >preselecciones Eugen Scholz, Riedel de Haën, Seelze, Ger- acondición: sí many indicar deriva: sí METROHM Application Bulletin No. 77: corr.deriva: Karl Fischer Water Determinations llamada ident: llamada peso: todo límites peso: horno: puls.activación: ------------ 758 KFD Titrino 758.0022...
  • Página 188 6.3 Métodos de usuario 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 18:06 Ipol >fórmula Water=EP1*C39*C01/C00/C02 RS1 texto Water RS1 num.decimales RS1 unidad: RS1 control val.límite: Titer=C39 RS2 texto Titer RS2 num.decimales RS2 unidad: mg/ml RS2 control val.límite: >cálculo silo igualar a Id: >variables comunes...
  • Página 189: Titraciones Kf Con Diferentes Reactivos Kf

    Si se utilizan reactivos diferentes al mismo tiempo hay que asignar los valores de los títu- los de los diferentes reactivos a diferentes variables comunes. Estos se incluyen en los cálculos de los métodos de titración correspondientes. 758 KFD Titrino...
  • Página 190: Determinación Del Título De Sosa Cáustica "Tit.naoh

    25.0 °C disuélvalo en 40 mL de agua destilada sin interv.tiempo carbonato. >condiciones de parada parada V: abs. Bibliografía: parada V 99.99 ml METROHM Application Bulletin No. 206: veloc.rell. max. ml/min Titer determinations in potentiometry >estadística status: sí media tab.res: original >preselecciones...
  • Página 191: Determinación Del Título De Ácido "Tit.hcl

    25.0 °C disuélvalo en 40 mL de agua destilada. interv.tiempo >condiciones de parada Bibliografía: parada V: abs. parada V 99.99 ml METROHM Application Bulletin No. 206: veloc.rell. max. ml/min Titer determinations in potentiometry >estadística status: sí media tab.res: original >preselecciones...
  • Página 192: Determinación De Los Valores P Y M "P+M Val

    6.3 Métodos de usuario 6.3.10 Determinación de los valores p y m "p+m val." Determinación de la capacidad de ácido de 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 14:47 agua y agua residuales. p+m val. parameters El valor p es la cantidad de ácido consumido >SET1...
  • Página 193 6.3 Métodos de usuario 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 14:48 p+m val. >fórmula p value=EP1*C36 RS1 texto p value RS1 num.decimales RS1 unidad: mmol/l RS1 control val.límite: m value=EP2*C36 RS2 texto m value RS2 num.decimales RS2 unidad: mmol/l RS2 control val.límite:...
  • Página 194: Determinación Del Índice De Bromo "Brnumber

    RS1 unidad: Bibliografía: RS1 control val.límite: >cálculo silo ASTM D 1159-84 igualar a Id: Metrohm Application Bulletin No.. 177 >variables comunes Nota: >impresión Los cambios en el potenical pueden ser impresión COM1:compl; bruscos. Por eso es mejor utilizar el modo >media...
  • Página 195: Determinación Del Índice De Formol "Formolno

    FormolNo RS1 num.deciamles RS1 unidad: RS1 control val.límite: >cálculo silo igualar a id: Bibliografía: >variables comunes Metrohm Application Bulletin No. 180 >impresión Automatic determination of the formol number impresión COM1:compl; in fruit and vegetable juices. >media MN1=RS1 >variables transitorias ------------...
  • Página 196 6.3 Métodos de usuario Preparación de los submétodos y adición de formaldehido "FormoPre" 758 KFD Titrino 758.0022 fecha 2001-11-21 hora 18:08 FormoPre En la primera titración SET se titra la muestra parameters a pH 8.5 con NaOH. >SET1 punto final EP pH 8.50...
  • Página 197 6.3 Métodos de usuario Submétodo de la determinación del índice de "FormoDet" Segunda titración SET con NaOH a pH 8.5 758 KFD Titrino 758.0022 después de la adición del formaldehido. fecha 2001-11-21 hora 18:10 FormoDet parameters >SET1 punto final EP pH 8.50...
  • Página 198: Validación Del Titrino

    Procedure) en los Boletines de aplicaciones Metrohm No. 252: Validation of Metrohm Titrators (potentiometric) according to GLP/ISO 9001. No. 255: Validation of Metrohm KF Titrators and KF Oven according to GLP/ISO 9001. El intervalo de validación se puede comprobar con el Titrino (ajuste bajo <CONFIG>, control).
  • Página 199: Garantía Y Conformidad

    6.5 Garantía y conformidad 6.5 Garantía y conformidad 6.5.1 Garantía La garantía METROHM cubre cualquier defecto de fabricación o material que pueda tener el aparato durante 12 meses, a partir de la fecha de entrega, y se reparará gratuitamente en nuestro taller. Sólo el transporte correrá...
  • Página 200: Declaración De Conformidad Ad Ue

    6.5 Garantía y conformidad 6.5.2 Declaración de Conformidad UE Declaración de Conformidad UE La sociedad Metrohm SA, Herisau, Suiza, certifica por la presente que el equi- 758 KFD Titrino cumple los requisitos de las directivas 89/336/CEE y 73/23/CEE de la UE.
  • Página 201: Certificado Para La Conformidad Y Validación Del Sistema

    The features of the system software are docu- mented in the instruction manual. Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servic- ing.
  • Página 202: Referencias De Pedido, Accesorios

    2 aparatos..................... 6.6008.500 1 Instrucciones para el uso del KFD Titrino 758 ..........8.758.1205 1 Instrucciones breves para el uso del KFD Titrino 758 ........8.758.1215 1 Guía de referencias rápidas del KFD Titrino 758 ..........8.758.1225 758 KFD Titrino...
  • Página 203 703 Ti Stand para titraciones KF ................2.703.0010 728 Agitador magnético ..................2.728.0040 727 Ti Stand, para aclarar y adición de disolvente nuevo ........2.727.0010 802 Agitador varilla ....................2.802.0010 727 Ti Stand con agitador magnético incorporado..........2.727.0100 758 KFD Titrino...
  • Página 204 Cable Titrino — Impresora EPSON LX300+ (9/25 polos) ........6.2134.050 Cable Titrino — HP Desk Jet (interfase en serie) (9/25 polos) ......6.2134.050 Cable Titrino — HP Desk/Laser Jet (IF en paralelo) ........... 2.145.0330 Para la conexión de impresora/balanza en el mismo COM ....6.2125.010+6.2125.030 758 KFD Titrino...
  • Página 205 Dosimat 665, 725, 765, 776..........6.2141.040 Cable Titrino — Cambiador de muestras 730, 760+ 2x Dosimat 665, 725, 765, 776 ........... 6.2141.050 Cable Titrino 758 KFD — Control Unit 664 del Cambiador de muestras 673/674 ........3.980.3560 758 KFD Titrino...
  • Página 206: Indice

    Límites del valor del resultado .... 43 CAL............38 Control remoto Cálculos ............ 42 vía "Remote"........173 Cálculos estadísticos ........ 45 vía RS232..........71ff ........66, 67 ........20, 30 cálculo silo corr.deriva: <CAL.DATA> ........6, 61 ........18, 28 crit.parada: 758 KFD Titrino...
  • Página 207 Identificación ........62, 64 DX sobrecargado llamada ......21, 30, 37, 40 ........67-68 igualar a id: Impresión ..........48ff ..........64 editar silo ..........12 impresión Electrodo, conexión.........167 Impresiones automáticas ......48 Emisión de datos ........48 ........48 impresión COMX: <ENTER>...........6 758 KFD Titrino...
  • Página 208 ......21, 31, 37, 40 llamar ident: Decimales..........43 Fabricación..........5 Muestra..........11 Malfunciones ......... 77ff, 137ff Pedido ..........198 Manejo manual ......... 6ff Serie............5 ......138 más de 500 puntos med. MEAS............32 Ordenes por control RS232 758 KFD Titrino...
  • Página 209 Silo punto final EP... Informes ......... 48, 68 Memoria..........63ff ..........140 silo lleno <QUIT> ............6 ..........140 silo vacío <SMPL DATA> ........6, 62ff Sondas, conexión ........167 RAM, inicializar ........154 Software Handshake....... 130 ............59 recargar 758 KFD Titrino...
  • Página 210 ......19, 29 tiempo extracción ........18, 28 tiempo parada ....... 141 tiempo parada alcanzado ......141 time out teclado PC ......... 141 TIP terminado TIP ............. 40, 52ff Titración EP............27 Problemas........26, 35 758 KFD Titrino...

Tabla de contenido