Braun ThermoScan IRT3030LA Manual De Instrucciones página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

All manuals and user guides at all-guides.com
Situation
Do you have any further
questions?
Ambient
temperature is not
within the allowed
operating range
(10 °C - 40 °C).
HI = too high
LO = too low
Temperature taken
is not within typical
human temperature
range (34°C - 42.2 °C)
System error
Battery is low.
Battery is too low.
Device cannot be
powered on to the
ready state.
Solution
Allow the thermometer to remain
for 30 minutes in a room where
the temperature is between
10 °C and 40 °C .
Make sure the probe tip and lens
are clean and a new, clean,
disposable lens filter is attached.
Make sure the thermometer is
properly inserted. Then, take a
new temperature.
Reset the thermometer
by removing the battery and then
replacing it again, wait 1 minute,
then turn on again.
Please contact your authorized
Service Center (See Guarantee
Card).
Insert new battery.
Insert new battery.
Please contact your authorized
Service Center (See Guarantee
Card).
English
7

Publicidad

Tabla de contenido
loading

Tabla de contenido