1. Put two clamp halves (one with threads and one without threads)
around the wall mount bracket. Refer to
2. Connect the clamp halves together with four bolts. Tighten the bolts
sufficiently to temporarily hold the clamp in position.
3. Put the other two clamp halves around the front end of the frame.
Refer to
Figure
10.
Note: Typically, the front of the frame will point toward the wall. Refer to
Figure 10
and
Figure 14
on page 16. If flow conditions make it necessary to
point the sensor away from the wall, use the 12-inch spacer and put the two
clamp halves around the back end of the frame.
4. Connect the clamp halves together with four bolts. Tighten the bolts
sufficiently to temporarily hold the clamp in position.
All manuals and user guides at all-guides.com
Figure
10.
Figure 10 Assemble the clamps on the wall bracket and frame
1 Adjustable wall bracket
2 Clamp half, threaded
3 Clamp bolt, ¼–20 x 1 in.
4 Clamp half, not threaded
Install the frame on the wall
D A N G E R
Explosion hazard. Review the safety information in
precautions
on page 5 before entering a confined space.
5 Frame
6 Spacer
7 Wall mount bracket
Confined space