• To clear the memory of the relevant user memory, you
must first select a user memory. Start a request for ei-
ther average values or individual measured values and
press and hold the SET button for approx. 3 seconds.
All the values in the current user memory are deleted
after 3 brief beeps are output.
• You have the option to delete individual measured val-
ues by pressing the SET button when the measured
values are displayed after the measurement.
• To switch off, press the START/STOP button
• If you forget to switch off the device, it will switch off
automatically after 3 minutes.

8. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries. The cuff line
should not be bent sharply.

9. Error messages / troubleshooting

In case of faults, the
message appears in the display.
Error messages can occur when
1: Systolic pressure is not detected,
2: Diastolic pressure is not detected,
All manuals and user guides at all-guides.com
In the above cases, you must repeat the measurement. Make sure
that the cuff tube is properly inserted and that you do not move or
talk. Re-insert the batteries if necessary, or else replace them.

10. Specifications

Model no.
Measurement
.
method
Measurement
range
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible oper-
ating conditions
26
3: The cuff is fastened too tightly or loosely,
4: The pump pressure is higher than 300 mmHg,
5: There is a system or unit error.
BM 40
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 –180 beats/minute
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 119 mm x W 109 mm x H 60 mm
Approx. 257 g (without batteries)
22 to 35 cm
+ 10 °C to + 40 °C, 15-93 % relative air hu-
midity (non-condensing)