3. Will not power up.
4. False alarms - the
TEK-Mate alarms
when the probe is
moved or bumped.
Return Authorization Procedure
All defective TEK-Mates should be returned to your wholesaler for warranty
evaluation. If you have any questions, please contact INFICON at
+49 221 56788-660 (servicetools.europe@inficon.com).
NOTE: Do not return your defective unit directly to the factory without first
contacting your wholesaler.
3a. Batteries are worn
out.
3b. Batteries have
been improperly
installed.
4a. Sensor leads are
bent.
4b. Moisture was
absorbed by the
sensor during a long
period without use.
8
3a. Install a new set of
batteries. See page 4.
3b. Check battery installation
as shown in
Figure 1. auf
Seite
4.
4a. Remove the sensor and
inspect the leads. Straighten
the leads with needle nose
pliers, if necessary, and
reinstall the sensor.
4b. Run the TEK-Mate for at
least 20 minutes. The
absorption of moisture does
not affect the life or sensitivity
of the sensor.