3. Soak
• Prepare a non-enzymatic detergent according to the manufacturer's
recommendations (0.125 ounces per gallon of tap water at
35–40 °C)
.
3
• Fully immerse the device and use a syringe to inject any lumens and
mated surfaces with 50 mL of the detergent.
• Soak the device for a minimum of 15 minutes.
4. Brush
• Thoroughly brush the exterior of the device using a soft-bristled
brush.
• Using a syringe, inject 50 mL of the detergent into any cannulae,
lumens, or mated surfaces a minimum of five times.
• Brush any lumens a minimum of five times from each end, using an
appropriate bottle brush.
• Actuate the device, brushing around any movable parts in all
extreme positions.
5. Rinse
• Thoroughly rinse the device with RO/DI water until all detergent
residue is removed. Flush any lumens or crevices a minimum of five
times. Once all detergent residue is removed, continue to rinse for a
minimum of 30 seconds.
• Drain the excess water from the device and dry it using a clean cloth
or pressurized air.
Prolystica® 2x Neutral Detergent is validated for cleaning efficacy.
3
Cleaning: Automated
1. Brush
• Using a syringe, inject 50 mL of the enzymatic detergent (from
the "Preparation for Cleaning" section) into any lumen and mated
surface a minimum of one time.
• Brush from both ends of any lumens a minimum of five times, using
an appropriate bottle brush.
2. Rinse
• Rinse the device with RO/DI water at ambient temperature
until there is no visible detergent residue. Continue to rinse for
a minimum of 30 seconds after all detergent residue has been
removed.
• Place the device in the washer on an incline to facilitate drainage.
EN-30