Aesculap Power Systems
Acculan® 3Ti charger GA677
Charging of Acculan® 3Ti rechargeable bat-
teries GA666 and GA676
Destruction of the product, charging bay
or battery due to inserting the wrong
type of battery in the charging bay!
CAUTION
Insert batteries only in their appro-
priate charging bays.
Insert batteries in their correct posi-
tion.
Fig. 1
Legend
A Mark on rechargeable battery
B Mark on battery compartment
Insert the battery with the contacts downwards in
the battery compartment, making sure that the
mark on rechargeable battery A is aligned with the
mark on battery compartment B, see Fig. 1.
8
"Charging progress" indicator
Battery charging starts automatically.
"Charging progress" indicator
Charging has been completed; remove
the battery from the charging bay.
The charging time depends on the
charge state and capacity of the bat-
teries.
"Battery change recommended" indi-
cator
Symbol is illuminated in addition to
"Charging completed" indicator.
The battery failed the automatic bat-
tery check.
A
It is recommended to replace the bat-
tery, see chapter Troubleshooting list.
B
"Action request" indicator
Charging cannot be performed nor-
mally; the "Charging progress" indica-
tor goes dark.
Possible cause: bad contacts; exces-
sive
charging, see chapter Troubleshooting
list.
battery
temperature
during