Operation
Command
Etc., when all changes are done:
Move cursor out of data
field
Cursor disappears and new value is stored
Move to date
Change date same way as described above. When all
changes are done:
Leave setup menu
RETURN
Confirm or decline
YES/NO
changes
Return to main screen
ESC
2.7.2 Other Setup Options
Other options in the setup menu are listed below. Default
value is bold or in brackets:
•
Time format: 12 / 24 hours
•
Date format: dd.mm.yyyy, mm/dd/yyyy
•
Medium density: 0.10...5.00 kg/l (1.00)
•
Pressure unit: bar, mbar, kPa, hPa
in
wcwc
•
Pressure measurement mode: normal
measurement is updated every four seconds) or fast
(measurement is updated every second)
•
Temperature unit: °C or °F
•
Display brightness: 20...100% (60%)
•
Printer: TD600, Other
•
Search by: Cust. num., Cust. name
•
Language: English, Deutsch
•
Printer logo: allows up to six lines of text which are
printed at the beginning of each printout with the
optional pocket printer. Default is: (line 1): Resideo,
(line 2) VM242A, (line 3) BasicMes-2
3
Technical Data
Media
Medium:
Water or water-glycol mixture,
quality to VDI 2035
(up to 50 % Glycol)
Pressure values
Static (burst) pressure: max. 32 bar (464 psi)
Differential pressure:
min. 0.05 bar (0.73 psi)
max. 17 bar (247 psi)
Differential pressure
up to 1 bar: 0.1 mbar
resolution:
above 1 bar: 1 mbar
Operating temperatures
Max. operating
-20 - 120 °C (-4 - 248 °F)
temperature medium:
GB
Key
ON
Æ
DOWN
È
ESC
ON/ESC
ESC
, m H
O, psi,
hPa
22
(i.e.
normal
Max. ambient
temperature:
Storage temperature:
Specifications
Accuracy:
Scan rate:
To guarantee perfect
functioning, a filter must
be inserted ahead of the
pressure reducing valve
Weight:
Dimensions:
Units and Display
Pressure units:
Temperature units:
Flow display:
k
-value input:
v
Density correction
factor:
Internal Temperature Measurement
Measuring range:
Accuracy:
Resolution:
External Temperature Measurement
Measuring range:
Accuracy:
Resolution:
4
5 - 40 °C (41 - 104 °F)
-20 - 60 °C (-4 - 140 °F)*
<3 % of measured value in range
of ±10 mbar better than ±0.3 mbar
at stable conditions in ambient
temperature range
Normal: 1Hz, average value of four
scans (time slot four seconds)
Fast: 4Hz, no averaging
USB, HP-IR for communication
with pocket printer
2.5 kg including accessories and
carry case
470 x 370 x 110 mm
(length x width x height)
mbar, bar, hPa, kPa, mH
O,
2
inH
O, psi
2
°C, °F
up to 5 mbar: no display
0 - 1.000 l/h: resolution 1 l/h
3
1 - 100 m
/h:
3
resolution 0.01 m
/h
3
100 - 1000 m
/h:
3
resolution 0.1 m
/h
3
1000 - 10000 m
/h: resolution
3
1 m
/h
k
< 1 in increments of 0.001
v
k
1...99.99 in increments of 0.01
v
k
100...999.9 in increments of 0.1
v
k
1000...10000 in increments of 1
v
0.10 - 5.00 kg/l in increments of
0.01 kg/l
-20 - 60 °C (-4 - 140 °F)
< ±1 K
0.1 °C
-20 - 300 °C (-4 - 572 °F)
±2 K from 0 °C to 133 °C
otherwise 1.5 % of actual value,
according to EN 50379-2
0.1 °C