TROUBLESHOOTING

PROBLEM
Erratic and/or
continuous
beeping signal
Erratic and/or
continuous
beeping signal
Reduced
sensitivity
Reduced
sensitivity
Probe pump
runs but unit
does not beep
Unit will not
detect a
known leak
Unit will not
detect a
known leak
Unit will not
detect a
known leak
Unit will not
detect a
known leak
* Remember that a worn out tip, no matter how clean, is still worn out and
will result in poor unit performance.
** Battery condition can be judged by checking the power indicator light
on the front of the unit. Weak batteries may cause the light to "blink".
Good batteries will result in a steady light.
NOTE: It is best to measure the battery voltage, as indicated in the
maintenance section on page 8, before reaching a conclusion.
CAUSE
Dirty tip
or
protector
Low battery
condition
Tip is worn
out
Dirty tip or
protector
Batteries
reversed
Tip is worn
out
Low battery
condition
Turned on in
presence of
a leak
Dirty tip
or Protector
9
CORRECTION
Replace or *
clean tip
or protector
Check pwr light **
& voltage. Replace
batteries if needed
Replace tip
Replace or clean *
tip or protector
Remove and
correct
Replace tip
Check power light**
& voltage. Replace
batteries if needed
Reset unit in a
clean atmosphere
(see page 6)
Clean or *
Replace tip
or Protector