Getting Started - Pari 22F81 Instrucciones De Uso

Ocultar thumbs Ver también para 22F81:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ENGLISH
I. GETTING STARTED
1. Make sure that you have all components illustrated on the front cover. If you did not
receive all components, contact the dealer where you purchased your PARI Nebulizer.
2. Read these Instructions For Use. Make note of all warnings and safety instructions. If
you do not understand any portion of these Instructions For Use, please call PARI
Customer Service using the telephone numbers listed on the back page.
3. Your nebulizer should last from 6 months to 1 year with general use and we have
included a sticker set to remind you to replace your nebulizer. Fill out and mail the
warranty card.
4. It is recommended that you buy a back-up nebulizer in case your nebulizer is broken
or have parts lost.
II. SET-UP
The PARI Nebulizer was designed to be used with a PARI compressor or with the central
air supply in a hospital. In order to achieve the specified droplet characteristics with the
central air supply, a liter flow of 3.5-8 liters per minute must be selected.
1. Unwind the tubing. For non-PARI compressors, go to step two. For use with the
PARI PRONEB
is included. If using these compressors, insert the tubing adapter into one end of the
tubing by pushing in firmly (Fig.A), the rigid side of the adapter.
2. Insert one end of the tubing directly to the compressor air outlet. Make sure the tubing
is firmly inserted or attached (Fig.B).
3. Make sure nebulizer has been cleaned and/or disinfected before use and since last
treatment (see opposite page for proper instructions). Remove the top of nebulizer by
twisting counter-clockwise until it stops and lift up (Fig.C).
4. Pour prescribed medication into bottom of nebulizer (Fig.D). Do not fill past the
max line. Caution: LC Star (part #22F51) is for 5 cc (ml) of medication maximum.
5. Replace the top of the nebulizer. The semi-circle should face the nebulizer outlet.
6. If using a mask (optional, part#'s 14F15 or 14F20), attach mask directly to the nebulizer
outlet (Fig.E).
7. If using the mouthpiece, attach the mouthpiece to the nebulizer outlet with the
valve facing up. Then attach the inspiratory valve by pressing firmly into the top of
the nebulizer. The valve cap is designed to fit snugly.
8. Connect the other end of the tubing to the air intake on the bottom of the nebulizer.
Press on firmly. The tubing is also designed to fit snugly on the air intake.
WARNING
!
Check PARI Nebulizer for proper assembly. All parts should be seated firmly in
place. Use of an improperly assembled nebulizer could prevent adequate delivery of
medication.
6
TM
, PRONEB TURBO
TM
, and PARI MASTER
®
compressors, a tubing adapter