OPERATING INSTRUCTIONS
Check the unit for proper operation before each use.
This unit has been designed for intermittent use. Normal
treatment time should be 5 to 10 minutes. Turn off the unit
after 20 minutes of use, and allow unit to cool down 40 minutes
before turning it on again.
1. As soon as you have unpacked the device check it for visible
damages or defects.
2. Check that accessories are complete and not damaged.
3. Before using the device, proceed with the cleaning operations
as described in the "CLEANING AND MAINTENANCE"
section.
4. Remove the lid from the main unit body (pic. 11).
5. Make sure that the nebulizer top is in the "Stand-by" position
(pic. 12), i.e.:
• the nebulizer top is seated snugly on the nebulizer
body and it is properly positioned in the nebulizer seat;
• the arrow indicators between the nebulizer top and the
nebulizer seat are aligned.
6. Pull the nebulizer kit out from the "stand-by" position
(pic. 13).
7. Open the nebulizer by pulling the top up (pic. 13).
8. Make sure that the micronization cone (blue color) is
correctly positioned inside the nebulizer chamber (pic. 15).
9. Pour the prescribed medication into the nebulizer body (pic.
18) as instructed by your doctor, paying attention not to
exceed the maximum allowed quantity (7 ml.).
Caution: Never pour medication into the nebulizer seat
(pump) area of the compressor. Pump damage will occur.
Place medication in the nebulizer cup only.