1
1
prepare Zero
press ZERO
1
2
3
4
Zero accepted
prepare Test
press TEST
Countdown
3:00
07/2017
Chlorine MR, total
3
with Powder Pack
0.02 – 3.5 mg/l Cl
1. Fill a clean vial (24 mm Ø) with 10 ml of the water
sample, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the
3. Press ZERO key.
4. Remove the vial from the sample chamber.
5. Add the contents of one VARIO Chlorine TOTAL-DPD /
F10 Powder Pack (blue color marking
from the foil to the water sample.
6. Close the vial tightly with the cap and swirl several times
to mix the contents (approx. 20 seconds).
7. Place the vial in the sample chamber making sure that
the
8. Press TEST key.
Wait for a reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l total
Chlorine.
Notes:
See notes Chlorine
2
marks are aligned.
marks are aligned.
) straight
41