Descargar Imprimir esta página

Lovibond Tintometer Manual Del Usuario página 48

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 128
1
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
48
COD HR
2
with Vario Tube Test
0.2 – 15 g/l O
Insert the adapter for 16 mm Ø vials.
1. Open one white capped reaction vial and add 0.2 ml
deionised water (this is the blank (Note 1)).
2. Open another white capped reaction vial and add
0.2 ml of the water sample (this is the sample).
3. Close the vials with the cap tightly. Invert the vial gently
several times to mix the contents.
(CAUTION: The vial will become hot during mixing!)
4. Heat the vials for 120 minutes in the preheated reactor
at a temperature of 150°C.
5. (CAUTION: The vials are hot!)
Remove the tubes from the heating block and allow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
6. Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are l aligned.
7. Press ZERO key.
8. Remove the vial from the sample chamber.
9. Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are l aligned.
10. Press TEST key.
The result is shown in the display in g/l COD.
^
( =
200 – 15 000 mg/l O
2
)
2
07/2017

Publicidad

loading