Overview
Structure

Recommended fields of application

Cleaning
Outside
cleaning
Aging of the
conductivity
measuring cell
4

EN English

1
2
3
1
Measuring electrode
2
Temperature sensor in graphite enclosure
3
Shaft
4
Closing head
 On site measurements in rivers, lakes and wastewater
 Fish farming
 Ground water measurements
 Applications in water laboratories
CAUTION
To clean the sensor, disconnect it from the instrument.
We recommend to clean the sensor thoroughly, especially before measuring low
conductivity values.
Contamination
Cleaning procedure
Lime sediments
Immerse in acetic acid for 5 minutes (volume share =
10 %)
Fat/oil
Clean with warm water containing washing-up liquid
After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.
Normally, the conductivity measuring cell does not age. Special measuring
mediums (e.g. strong acids and bases, organic solvents) or temperatures that
are too high may considerably reduce its lifetime or lead to damage. The
warranty does not cover cases where such conditions cause failure or mecha-
nical damage.
4
ba75841y04
CO 11
05/2021