Troubleshooting; Motor Not Running; Pump Is Too Loud; Insufficient Vacuum - Medela Clario Toni Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 258

13 Troubleshooting

Motor not running

Check if:
– the Clario Toni is switched on�
– the mains voltage is correct and if the mains plug is inserted correctly into the fixed mains
socket�
– the internal battery is charged�

Pump is too loud

Check the position and seal of the valve discs inside the SafetyChamber�

Insufficient vacuum

Check if:
– the vacuum regulator is set correctly�
– the tubings are not defectiv e or broken� If necessary, replace�
– all plug-in connections are tight�
– the suction canister / lid have no cracks, brittle areas, discolouration� If necessary, replace�
– the seal of the valve discs inside the SafetyChamber is tight�
– the membrane is clean and has no tears�

Running time from battery insufficient

Calibrate the rechargeable battery� See chapter 12�

No LED lit

The rechargeable battery is fully charged�

LED lit up

– The rechargeable battery is being charged�
– Pump is plugged in�

LED blinks slowly

The rechargeable battery is soon empty� The battery power is sufficient for about 10 minutes
of pumping�

LED blinks rapidly

The rechargeable battery is defective� For replacing the defective battery see chapter 10�
24 |
Troubleshooting

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido