ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacturer's declaration – electromagnetic emissions
The IH21 is intended for use in the electromagnetic environment specified below. The customer or the user
of the IH21 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The IH21is intended for use in the electromagnetic environment specified below. The customer or the user of
the IH21should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Power frequency (50Hz)
magnetic field
IEC 61000-4-8
Compliance Electromagnetic environment – guidance
Group 1
The IH21 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class B
The IH21 is suitable for use in all establishments, inclu-
ding domestic establishments and those directly con-
nected to the public low-voltage power supply network
Class A
that supplies buildings used for domestic purposes.
Complies
IEC 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3 A/m
3 A/m
70
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30 %.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.