Troubleshooting - FLAEM AlphaNeb Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43

TROUBLESHOOTING

Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
The device does not work
The device does not
nebulize or
nebulizes insufficiently.
The apparatus is noisier
than usual
If after verifying the above mentioned conditions the device should not operate
properly, contact your trusted retailer or the nearest authorized FLAEM service
center.
CAUSE
The power cable has not
been correctly inserted into
the power socket
The medication has not
been inserted in the
nebulizer
The nebulizer has not been
properly fitted.
The nebulizer nozzle is
clogged
The air pipe is not correctly
connected to the device
The air pipe is bent, dam-
aged or twisted
The air filter is dirty
The Filter is not correctly
inserted in its housing
14
SOLUTION
Correctly insert the power
cable in sockets.
Pour the right amount of
medication in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as
shown in the connection
diagram on the cover.
Disassemble the nebulizer,
remove the nozzle and perform
cleaning operations Failure to
remove medication deposits
from the nebulizer affects its
efficiency and operation.
Strictly comply with the
instructions contained in the
"CLEANING, SANITISATION,
DISINFECTION" chapter
Ensure the device's accessories
are properly connected to
the air intake (see connection
diagram on the cover).
Unwind the pipe and ensure it
isnot crushed or punctured. If
necessary replace it.
Replace the filter
Fully and properly insert the
Filter in its housing

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

P0707em f400

Tabla de contenido