Sunoptic Surgical Titan x450 Manual De Uso página 11

Ocultar thumbs Ver también para Titan x450:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 88
6.2.1.1
(Refer to 8.1 LAMP REPLACEMENT) Disconnect power. Open the door. The lamp
positioning lever should be completely down(horizontal) to properly engage the lamp
and completely up (vertical) to release it. Raise the lever completely if not already
raised and firmly push the lamp to assure full engagement to the power connectors.
Lower the lamp engagement lever counter-clockwise, from vertical to horizontal, and
observe that the lamp rises slightly as it is engaged into proper operating position.
6.2.1.2
6.2.1.3
The Tx450D digital control system communicates with Sunoptic Surgical lamps designed
for it. Only Sunoptic Surgical lamps can be engaged by the positioning mechanism.
LIT-219 Sunoptic Surgical
Rev. C
ENGAGE LAMP
CLOSE DOOR
Close the door completely. (Fan will continue to run with door open.)
INCOMPATIBLE LAMP MODULE INSTALLED
Install correct SSX0450 lamp.
®
Page 11 of 169
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido