Sunoptic Surgical Titan x450 Manual De Uso página 12

Ocultar thumbs Ver también para Titan x450:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 88
6.2.1.4
Installed lamp does not match the intended power (watts). Tx450 units require a 400-watt lamp.
6.2.1.5
Lamp has been used more than 1000 hours. Install new SSX0450 lamp or one used
less than 1000 hours.The main screen always displays lamp hours, and notices are
displayed when a lamp is approaching the end of its useful lamp life. This screen is
displayed when the lamp has exceeded its usable lifetime.
6.2.1.6
In the event of a fan problem the unit will turn off the lamp and display this screen.
Turn off and restart the illuminator. If the fault persists the unit needs servicing
LIT-219 Sunoptic Surgical
Rev. C
INCORRECT WATTAGE LAMP INSTALLED
Install correct SSX0450 lamp.
LAMP HAS EXCEEDED USEABLE LIFE
FAN FAULT
®
Page 12 of 169
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido