Connection A - High Vacuum Flange; Connection Configurations - Varian 969-9369 Manual De Instrucciones

Ocultar thumbs Ver también para 969-9369:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12

Connection A - HIGH VACUUM FLANGE

To connect the Turbo pump to the ISO inlet flange,
remove the outer ring and position the centering
ring as shown in the figure.
Then fix the two flanges with the clamps or claws
as shown in the figure.

Connection configurations

All manuals and user guides at all-guides.com
33
TECHNICAL INFORMATION
For ConFlat flange connections we recommend
using Varian hardware.
To facilitate assembly and dismantling, apply Fel-
pro C-100 high temperature lubricant to the screw
threads protruding from the flange and between
the nuts and flange.
Attach the units and tighten each one in turn. Re-
peat the sequential tightening until the flange faces
meet.
! ! ! !
CAUTION
Exercise care when tightening nuts and bolts to
avoid creating dents in the envelope as this may
cause the pump rotor to lock.
87-900-891-01(C)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969-9370969-9371969-9372

Tabla de contenido