Guidance and Manufacturer's Declaration – Emissions all Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of this device
should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
Class B
RF Emissions
Radiated and Conducted
CISPR 11
Emissions
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC 61000-3-3
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
Transient/burst
±2kV on AC Mains
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
>95% Dip for
Voltage dips, short
0.5 Cycle
interruptions and
60% Dip for
voltage variations
5 Cycles
on power supply
30% Dip for
input lines
25 Cycles
IEC 61000-4-11
>95% Dip for
5 Seconds
Power Frequency
50/60Hz Magnetic
Field IEC 61000-4-8
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
SE-DV56-1
3A/m
3 Vrms from
Electromagnetic Enforcement – Guidance
This device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
This device is suitable for use in all establishments including domestic,
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance Level
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
±6kV contact
tile. If floors are synthetic, the relative humidity
±8kV air
should be at least 30%.
Mains power quality should be that of a typical
±2kV on AC Mains
commercial or hospital environment.
±1kV Differential
Mains power quality should be that of a typical
±2kV Common
commercial or hospital environment.
>95% Dip for
0.5 Cycle
Mains power quality should be that of a typical
60% Dip for
commercial or hospital environment. If the user
5 Cycles
of this device requires continued operation
30% Dip for
during power mains interruptions, it is
25 Cycles
recommended that the device be powered from
>95% Dip for
an uninterruptible power supply or battery.
5 Seconds
Power frequency magnetic fields should be that
3A/m
of a typical location in a typical commercial or
hospital environment.
Portable and mobile RF communications
equipment should be separated from the device
by no less than the recommended separation
V1 = 3 Vrms
distances calculated/listed below:
D=(3.5/V1)√ P
EN - 15