Preventive health monitoring

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Harmonic current emission
EN
Voltage fluctuations and flicker
Immunity test levels
Test
Electrostatic discharge
(IEC 61000-4-2)
Radiated RF EM fields
(IEC 61000-4-3)
Proximity fields from RF wireless
communications equipment
(IEC 61000-4-3)
Electrical fast transients and
bursts (IEC 61000-4-4)
Surge Immunity (IEC 61000-4-5)
Conducted disturbances induced
by RF fields (IEC 61000-4-6)
Rated power frequency magnetic
fields (IEC 61000-4-8)
14
IEC 61000-3-2, Class A, power < 75W
IEC 61000-3-3
Test levels
Contact Discharge: ±8 kV
Air Discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV
80-2700 MHz; 1kHz AM 80 %; 10 V/m
385 MHz; Pulse Modulation: 18 Hz; 27 V/m
450 MHz; FM: ±5 kHz Hub, 1 kHz Sinus; 28 V/m
710, 745, 780 MHz; Pulse Modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse Modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse Modulation: 217 Hz; 28 V/m
2450 MHz; Pulse Modulation: 217 Hz; 28 V/m;
5240, 5500, 5785 MHz; Pulse Modulation: 217 Hz; 9 V/m
Power supply lines: ±2 kV, 100 kHz repetition freq.
Signal / Data lines: ±1 kV, 100 kHz repetition freq.
0.5, 1 kV; Differential mode
Surge Waveform: Combination Wave
(2 µs x 50 µs Voltage, 8 µs x 20 µs Current)
0.15-80 MHz; 1 kHz AM 80 %; 3 Vrms, 6 Vrms ISM band
30 A/m, 50 Hz / 60 Hz
Table N°5: EMI test levels
4-3-6-2 Bloom Lab IFU, 09.04.2020, V1.0