ResMed Astral Guia Del Usuario página 11

Batería externa
Ocultar thumbs Ver también para Astral:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 306
Battery output
current (rated)
Protection
Dimensions
(L x W x H)
Battery weight
Typical life cycle
Operating conditions
Charging:
Discharging:
Transport/storage
conditions
Operating/storage air
pressure
Recharge time
Electromagnetic
compatibility
Aircraft use
IEC 60601-1
classification
Battery run time
Operating hours with a new battery under normal conditions. Test
1
conditions: Adult, P(A)CV mode, P insp: 20 hPa, PEEP: Off, Rate: 15 BPM,
Ti: 1.2s. Time may vary with environmental conditions.
Note: The manufacturer reserves the right to change these
specifications without notice.
Symbols
The following symbols may appear on the device or packaging:
 Attention, consult accompanying documents;
instructions before use;
objects and against vertically dripping water;
equipment;
number;
4.5 A / 3.75 A
Overcharge, overdischarge, excess current,
short circuit, high temperature
168 mm x 104 mm x 72 mm
1.2 kg typical
400 charge cycles at 23°C to 80% capacity
10°C to 35°C; 5–93% maximum humidity
0°C to 40°C; 5–93% maximum humidity
-20°C to +50°C; 5–93% relative humidity (non-
condensing)
700 hPa to 1100 hPa
< 6 hrs to full level (100%)
Product complies with all applicable
electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for
residential, commercial, and light industry
environments. Information regarding the
electromagnetic emissions and immunity of
these ResMed devices can be found on
www.resmed.com/downloads/devices.
ResMed confirms that the External Battery
meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
IATA classification for internal battery: UN 3481
– Lithium-ion batteries contained in equipment.
External Battery is UN3480
Class II (double insulation) and/or Internally
Powered Equipment, IP21, Continuous
Operation (from mains), Limited Operation
(from battery), Equipment not suitable for use
in the presence of a flammable anaesthetic
mixture with air, or nitrous oxide.
The External Battery, while connected to a
device and discharging, is rated to IP21 (Drip
proof) according to IEC60529. The AC power
supply and the Astral External Battery are both
rated to IP21.
< 8 hrs of typical use
 Protected against finger sized
21
 European RoHS;
 Catalogue number;
1
 Class II
 Batch Code;
8
 Humidity limitation;
 Read
 Serial
English
9