4. Maintenance
Refer to the Maintenance Guide OM-Z0135E.
This handpiece can be cleaned and disinfected with a Thermo-Disinfector.
5. Cleaning the Cellular Glass Optic Rod ( X10L,X12L,X15L,X20L,X25L,X85L,X95L )
Wipe clean the Cellular Glass Optic Rod entry point and exit point with an alcohol-immersed cotton swab. Remove all debris and oil. ( Fig.4 )
CAUTION
Do not use a sharp tool to clean the Cellular Glass Optic Rod. It could damage the glass and reduce
the light transmission. If illumination becomes dim contact your dealer.
This handpiece can be autoclavable up to Max.135°C.
X15L
Cotton Swab
Glass Rod End
Fig. 4